: This study compares system-reported organ doses (ODs) to manually calculated mean glandular doses (MGDs) in mammography across multiple centers and manufacturers in Dubai. : A retrospective study of 2754 anonymized mammograms from six clinics in Dubai were randomly retrieved from a central dose survey database. Organ doses were documented along with other dosimetry information like kVp, mAs, filter, target, compression force, and breast thickness. Mean glandular doses, MGDs, were calculated manually for all the patients using the Dance formula and inferential statistical analyses were run to compare the two figures and verify the factors affecting each. : Our study's analysis revealed that manually calculated mean glandular doses (MGDs) provide a more reliable indicator of radiation exposure than organ doses (ODs) reported by DICOM, particularly in multi-vendor scenarios. Manually calculated MGD values were consistently lower than system-reported ODs (MLO view: 0.96 ± 0.37 mGy vs. 1.38 ± 0.45 mGy; CC view: 0.81 ± 0.33 mGy vs. 1.22 ± 0.38 mGy). Significant differences in both system-reported ODs and manually calculated MGDs were observed across centers ( < 0.001). Strong correlations between system-reported ODs and manually calculated MGDs were found for Siemens equipment (r = 0.923, < 0.001) but only moderate correlations for GE systems (r = 0.638, < 0.001). Calculated MGD values were significantly higher for GE equipment compared to Siemens (1.49 ± 0.77 mGy vs. 0.93 ± 0.33 mGy, < 0.001). : This study addresses concerns regarding mammography dosimetry accuracy by demonstrating the superiority of mean glandular doses over DICOM-generated organ doses. These findings empower practitioners to optimize dose levels, ensuring safer and more effective breast cancer screening protocols.

Download full-text PDF

Source
http://dx.doi.org/10.3390/diagnostics15010081DOI Listing

Publication Analysis

Top Keywords

manually calculated
24
organ doses
16
glandular doses
16
calculated glandular
12
ods manually
12
doses mgds
12
system-reported ods
12
system-reported organ
8
calculated
8
doses
8

Similar Publications

: Prostate cancer (PCa) is the most frequent neoplasia in the male population. According to the International Society of Urological Pathology (ISUP), PCa can be divided into two major groups, based on their prognosis and treatment options. Multiparametric magnetic resonance imaging (mpMRI) holds a central role in PCa assessment; however, it does not have a one-to-one correspondence with the histopathological grading of tumors.

View Article and Find Full Text PDF

: This study compares system-reported organ doses (ODs) to manually calculated mean glandular doses (MGDs) in mammography across multiple centers and manufacturers in Dubai. : A retrospective study of 2754 anonymized mammograms from six clinics in Dubai were randomly retrieved from a central dose survey database. Organ doses were documented along with other dosimetry information like kVp, mAs, filter, target, compression force, and breast thickness.

View Article and Find Full Text PDF

Objective: This study aims to identify 18F-FDG-PET imaging features for improving treatment response evaluation in patients with necrotizing otitis externa (NOE), aiding in the difficult differentiation between sterile inflammation and active infection.

Study Design: Retrospective cohort study.

Setting: Tertiary hospital.

View Article and Find Full Text PDF

Impaired walking ability and leg health are commonly seen in broilers and can negatively impact their welfare. Commonly, walking ability and leg health are assessed manually, but this is time consuming and can be subjective. Automated approaches for scoring walking ability and leg health at the individual level could therefore have great added value.

View Article and Find Full Text PDF

Background: Glioblastoma is characterized by neovascularization and diffuse infiltration into the adjacent tissue. T2*-based dynamic susceptibility contrast (DSC) MR perfusion images provide useful measurements of the biomarkers associated with tumor perfusion. This study aimed to distinguish infiltrating tumors from vasogenic edema in glioblastomas using DSC-MR perfusion images.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!