The accuracy of measuring the activity of Lu is crucial for calibrating the instruments employed in clinical applications. To support the measurements of Lu solution, the existing TDCR-Čerenkov method has been extended for this nuclide. The contribution of β and γ transitions from Lu was taken into account in the extended method. Finally, the results obtained using this method, with an uncertainty of 0.62%, were found to be in good agreement with that derived from liquid scintillation counting.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.apradiso.2025.111660 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!