Accuracy and feasibility of continuous glucose monitoring system in pancreatectomy patients.

Langenbecks Arch Surg

Department of Surgery, Division of HBP Surgery & Liver Transplantation, Korea University College of Medicine, Seoul, Korea.

Published: January 2025

Purpose: Pancreatectomy patients often experience challenging fluctuations in blood glucose levels; therefore, they require a reliable monitoring system. This study aimed to determine the accuracy and acceptability of a continuous glucose monitoring (CGM) system compared with the intermittent capillary glucose test in patients who have undergone pancreatectomy.

Methods: Thirty non-diabetic pancreatectomy patients participated. We used the FreeStyle Libre Flash Glucose Monitoring System (Abbott Diabetes Care) for continuous interstitial glucose monitoring. Capillary reference glucose levels were checked four times daily. Accuracy was checked using the Clarke Error Grid.

Results: The mean age of the participants was 56.8 ± 12.0 years, of whom 61.3% underwent pancreaticoduodenectomy and 38.7% underwent distal pancreatectomy. Three patients developed pancreatogenic diabetes after pancreatectomy. The clinical accuracy of continuous glucose monitoring compared with capillary glucose was 43.9% in Zone A (clinically accurate zone) and 99.8% in Zone A + B (low risk of error) of the Clarke Error Grid. No device-related adverse events were reported. Patients rated favorable user acceptability on the questionnaire.

Conclusion: This pilot study demonstrated that the CGM device is accurate and safe for patients who underwent pancreatectomy, with favorable user acceptability. Despite these challenges, the study proposes that the CGM device is beneficial for monitoring glucose levels after discharge in patients with impaired glucose levels following pancreatectomy.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00423-024-03601-8DOI Listing

Publication Analysis

Top Keywords

glucose monitoring
20
glucose levels
16
continuous glucose
12
monitoring system
12
pancreatectomy patients
12
glucose
11
patients
8
capillary glucose
8
clarke error
8
favorable user
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!