Objectives: To evaluate the image quality and lung nodule detectability of ultralow-dose CT (ULDCT) with adaptive statistical iterative reconstruction-V (ASiR-V) post-processed using a deep learning image reconstruction (DLIR)-based image domain compared to low-dose CT (LDCT) and ULDCT without DLIR.
Materials And Methods: A total of 210 patients undergoing lung cancer screening underwent LDCT (mean ± SD, 0.81 ± 0.28 mSv) and ULDCT (0.17 ± 0.03 mSv) scans. ULDCT images were reconstructed with ASiR-V (ULDCT-ASiR-V) and post-processed using DLIR (ULDCT-DLIR). The quality of the three CT images was analyzed. Three radiologists detected and measured pulmonary nodules on all CT images, with LDCT results serving as references. Nodule conspicuity was assessed using a five-point Likert scale, followed by further statistical analyses.
Results: A total of 463 nodules were detected using LDCT. The image noise of ULDCT-DLIR decreased by 60% compared to that of ULDCT-ASiR-V and was lower than that of LDCT (p < 0.001). The subjective image quality scores for ULDCT-DLIR (4.4 [4.1, 4.6]) were also higher than those for ULDCT-ASiR-V (3.6 [3.1, 3.9]) (p < 0.001). The overall nodule detection rates for ULDCT-ASiR-V and ULDCT-DLIR were 82.1% (380/463) and 87.0% (403/463), respectively (p < 0.001). The percentage difference between diameters > 1 mm was 2.9% (ULDCT-ASiR-V vs. LDCT) and 0.5% (ULDCT-DLIR vs. LDCT) (p = 0.009). Scores of nodule imaging sharpness on ULDCT-DLIR (4.0 ± 0.68) were significantly higher than those on ULDCT-ASiR-V (3.2 ± 0.50) (p < 0.001).
Conclusion: DLIR-based image domain improves image quality, nodule detection rate, nodule imaging sharpness, and nodule measurement accuracy of ASiR-V on ULDCT.
Key Points: Question Deep learning post-processing is simple and cheap compared with raw data processing, but its performance is not clear on ultralow-dose CT. Findings Deep learning post-processing enhanced image quality and improved the nodule detection rate and accuracy of nodule measurement of ultralow-dose CT. Clinical relevance Deep learning post-processing improves the practicability of ultralow-dose CT and makes it possible for patients with less radiation exposure during lung cancer screening.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00330-024-11317-y | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!