Objectives: Careful consideration of the pre-analytical process for urine examination is essential to avoid errors and support accurate results and decision-making. Our objective was to assess the impact of various pre-analytical factors on urine test strip and quantitative chemistry results, including stability, tube type, fill volume, and centrifugation.
Methods: Residual random urine specimens were identified. Stability of 10 urine test strips and 13 quantitative chemistry parameters were assessed at eight time points (2, 4, 6, 8, 24, 48, 72, and 96 h) at room temperature (RT) and 2-8 °C (n=10-20 samples). The effect of additional pre-analytical variables was assessed, including using preservative tubes for urine chemistry as well as preservative tube underfilling and centrifugation on urine test strip results (n=10 samples).
Results: Seven of the ten urine tests strips evaluated met the minimal agreement criteria for stability (Cohen's kappa >0.70) across all conditions. A Cohen's kappa value of <0.70 was observed for pH (48 h), glucose (72 h), and protein (96 h) at RT. All 13 urine chemistry analytes remained stable at defined time points and conditions. Underfilling preservative tubes for urine test strips and centrifugation demonstrated no significant effect. The impact of using preservative tubes for urine chemistry was negligible with the exception of sodium and osmolality.
Conclusions: These findings highlight the pre-analytical factors that impact urine specimen evaluation and may be useful in informing clinical laboratory practices. Acceptable stability window for urine test strips should be considered in the context of the proportion of pathological samples evaluated.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/cclm-2024-1233 | DOI Listing |
BMC Med Inform Decis Mak
January 2025
Department of Pediatrics, School of Medicine, Ekbatan Hospital, Hamadan University of Medical Sciences, Hamadan, Iran.
Background: Urinary tract infection (UTI) is a frequent health-threatening condition. Early reliable diagnosis of UTI helps to prevent misuse or overuse of antibiotics and hence prevent antibiotic resistance. The gold standard for UTI diagnosis is urine culture which is a time-consuming and also an error prone method.
View Article and Find Full Text PDFBiol Trace Elem Res
January 2025
Clinical Research Centre, University Medical Centre Ljubljana, Zaloška Cesta 2, 1000, Ljubljana, Slovenia.
Premature ovarian insufficiency (POI) is poorly understood, with causes identified in only 25% of cases. Emerging evidence suggests links between trace elements (TEs) and POI. This study is the first to compare concentrations of manganese (Mn), copper (Cu), zinc (Zn), selenium (Se), molybdenum (Mo), arsenic (As), cadmium (Cd), mercury (Hg), and lead (Pb) across urine, serum, and whole blood in women with POI compared to healthy controls (HC), aiming to explore their distribution and potential associations with POI.
View Article and Find Full Text PDFAnal Chim Acta
January 2025
College of Chemistry, Beijing University of Chemical Technology, Beijing, 100029, China. Electronic address:
Background: The rapid and sensitive detection of nitrite is important to human health protection due to its carcinogenic and teratogenic risks with excessive intake. The Griess assay is widely applied for the design of nitrite detection system. However, its relatively slow reaction kinetics and sole colorimetry mode might limit it's the sensitivity and practical application.
View Article and Find Full Text PDFClin Chem Lab Med
January 2025
Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.
Objectives: Careful consideration of the pre-analytical process for urine examination is essential to avoid errors and support accurate results and decision-making. Our objective was to assess the impact of various pre-analytical factors on urine test strip and quantitative chemistry results, including stability, tube type, fill volume, and centrifugation.
Methods: Residual random urine specimens were identified.
Int J Med Sci
January 2025
Department of General Medicine, Yalamanchi Hospitals and Research Centre, Vijayawada-520002, Andhra Pradesh, India.
LN18178 is a standardized, synergistic combination of fruit rind and seed extracts, which has been reported to increase serum testosterone levels in young and aging males. The present 84-day randomized, double-blind, placebo-controlled study assessed the efficacy of LN18178 on the sexual function of aging male volunteers (age: 40-70 years; serum total testosterone: ≥ 300 ng/dL). The subjects with mild to moderate erectile dysfunction [5-item version of the International Index of Erectile Function (IIEF-5) scores 17-25] and low sexual desire (score < 3 on items 11 and 12 of IIEF) participated in this investigation.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!