Purpose: To assess the safety of acoustic radiation force optical coherence elastography in the crystalline lens in situ.

Methods: Acoustic radiation force (ARF) produced by an immersion single-element ultrasound transducer (nominal frequency = 3.5 MHz) was characterized using a needle hydrophone and used for optical coherence elastography (OCE) of the crystalline lens. Preamplified signals at 50, 100, 250, 500, 750, 1000, and 1250 mV peak amplitude were tested on ex vivo porcine eyes (n = 21). Three-dimensional optical coherence tomography (OCT) and confocal microscopy images were acquired before and after ARF exposure to each signal amplitude to determine damage.

Results: The acoustic intensity of the ultrasound transducer at 100-mV preamplified peak amplitude input demonstrated a signal-to-noise ratio high enough for tracking elastic wave propagation in the lens and spatial-peak pulse-average (SPPA) intensity of 24.1 W/cm² and mechanical index (MI) of 0.46. The SPPA intensity was lower than the U.S. Food and Drug Administration (FDA) safety limit (28 W/cm2), but the MI was twice the safety limit (0.23). OCT structural and confocal microscopy images showed damage only at levels exceeding 1150 W/cm2 and 3.2 for SPPA intensity and MI, respectively.

Conclusions: OCT and confocal microscopy showed that, even when the intensity exceeded FDA recommendations (>100 mV), no noticeable damage was observed. Although a further reduction in acoustic intensity is necessary to meet FDA safety limits, ARF-based elastography shows promise for safe clinical translation in quantitatively characterizing lenticular biomechanical properties.

Translational Relevance: This work assessed the safety standards for acoustic radiation force to be used in human lens elastography according to the FDA safety limits.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11684484PMC
http://dx.doi.org/10.1167/tvst.13.12.36DOI Listing

Publication Analysis

Top Keywords

acoustic radiation
16
radiation force
16
optical coherence
16
coherence elastography
12
crystalline lens
12
confocal microscopy
12
sppa intensity
12
fda safety
12
force optical
8
elastography crystalline
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!