Comparison of the CT-based micromotion analysis method versus marker-based RSA in measuring femoral head translation and evaluation of its intra- and interobserver reliability: a prospective agreement diagnostic study on 27 patients up to 1 year.

Acta Orthop

Department of Orthopaedics, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg, Gothenburg; Department of Orthopaedics, Sahlgrenska University Hospital, Region Västra Götaland, Gothenburg, Sweden.

Published: January 2025

Background And Purpose:  Computed tomography radiostereometric analysis (CT-RSA) assesses implant micromovements using low-dose CT scans. We aimed to investigate whether CT-RSA is comparable to marker-based radiostereometric analysis (RSA) measuring early femoral head migration in cemented stems. We hypothesized that CT-RSA is comparable to marker-based RSA in evaluating femoral head subsidence.

Methods:  We prospectively included 31 patients undergoing cemented total hip arthroplasty (THA), of which 27 were eligible for the analysis. Femoral head migration at 1 year was measured with marker-based RSA and CT-RSA. Comparison was performed using paired analysis and Bland-Altman plots, and the intra- and interobserver reliability of CT-RSA was assessed Results: The median (interquartile range [IQR]) translation on the Y-axis measured with marker-based RSA was -0.86 mm (-1.10 to -0.37) and -0.83 mm (-1.11 to -0.48) for CT-RSA (i.e. subsidence), with a median difference of -0.03 mm (95% confidence interval [CI] -0.08 to 0.18). The minimal important difference in translation was set to 0.2 mm. This value was excluded from the CI of the differences. No statistical difference was found between marker-based RSA and CT-RSA regarding assessment of subsidence of the femoral head. The Bland-Altman plots showed good agreement between the 2 methods in measuring subsidence of the femoral head. The intra- and interobserver reliability of the CT-RSA method was excellent with intraclass correlation coefficient (ICC) = 1 (0.99-1) and ICC = 0.99 (0.99-1), respectively.

Conclusion:  We showed that CT-RSA was comparable to marker-based RSA in measuring femoral head subsidence. Moreover, the intra- and interobserver reliability of the CT-RSA method was excellent, suggesting that the method is assessor independent.

Download full-text PDF

Source
http://dx.doi.org/10.2340/17453674.2024.42705DOI Listing

Publication Analysis

Top Keywords

femoral head
28
marker-based rsa
24
intra- interobserver
16
interobserver reliability
16
rsa measuring
12
ct-rsa comparable
12
comparable marker-based
12
reliability ct-rsa
12
ct-rsa
10
measuring femoral
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!