Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: This study aimed to assess the feasibility of implementing the Idylla system, an ultra-rapid, cartridge-based assay, as an extension of rapid on-site evaluation (ROSE) in cytology. The authors conducted a pilot validation study on specimens from non-small cell lung carcinoma, thyroid carcinoma, and melanoma, evaluating four assays designed to detect alterations in KRAS, EGFR, BRAF, gene fusions, and expression imbalances in ALK, ROS1, RET, NTRK1/2/3, and MET exon 14 skipping transcripts. They investigated the feasibility of providing accurate biomarker molecular testing results in a cytopathology laboratory within hours of specimen collection.
Methods: The authors evaluated the performance characteristics and turn-around-time of the Idylla system by testing a total of 144 cartridge assays across various specimen types, including fine-needle aspirate smears, formalin-fixed paraffin-embedded (FFPE) cell blocks, small tissue biopsy FFPE blocks, and control cell line FFPE scrolls.
Results: The average time from specimen input to results output was 2-3 hours. Accuracy across the four cartridge types was: KRAS assay: 100%, EGFR assay: 94%, BRAF assay: 100%, and GeneFusion assay: 94%. Analytical sensitivity ranged from 1% to 5% variant allele frequency for all assays. Inter-assay precision and analytical specificity were both 100%.
Conclusion: Using the Idylla system, actionable genetic alterations can be reliably detected within 2-3 hours from cytology and small biopsy samples with minimal input requirements. The findings of this study demonstrate the feasibility of incorporating same-day molecular testing as part of ROSE procedures in the cytopathology laboratory, ultimately shortening the time from procedure to personalized treatment for cancer patients.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/cncy.22930 | DOI Listing |