AI Article Synopsis

  • This study examined the effectiveness of bone SPECT imaging with and without SwiftScan technology in patients with prostate cancer bone metastases.
  • Involving 10 patients and 130 metastatic lesions, the research compared the quality of images produced at various acquisition times, showcasing improved quality scores and correlation coefficients with SwiftScan.
  • Findings indicated that SwiftScan allowed for a 50% reduction in acquisition time while still maintaining acceptable image quality and accurate quantitative measurements like SUV and contrast-to-noise ratios.

Article Abstract

Objectives: This study aimed to evaluate the quality and associated quantitative values of bone single-photon emission computed tomography (SPECT) with and without SwiftScan using a semiconductor camera equipped with a cadmium-zinc-telluride detector.

Methods: Ten patients with bone metastases from prostate cancer who underwent list-mode SPECT/computed tomography using a whole-body semiconductor camera participated in this study. A total of 130 metastatic lesions from 10 patients were analyzed. Standard SPECT images were obtained approximately 3 h later, and the images were constructed with and without SwiftScan.

Results: The visual assessment of 3-dimensional maximum intensity projection images showed that when an image quality score of 4 (good) or better was considered clinically acceptable, it was maintained at 4 or better in the 75% and 50% scans with SwiftScan, whereas only the 75% scan was considered acceptable without SwiftScan. The intraclass correlation coefficient was 0.952 at 5% for the standard time without SwiftScan and 0.990 with SwiftScan. The maximum standardized uptake value (SUV) changes were 0 to 9.5 (median 1.1) at 75%, 0.1 to 11.5 (1.65) at 50%, 0 to 15.7 (2.1) at 25%, 0.1 to 33.2 (4.2) at 10%, 0.2 to 8.9 (5.65) at 5% without SwiftScan. On the contrary, the SUV changes in absolute value were 0 to 5.4 (median 0.8) at 75%, 0 to 6.5 (1.4) at 50%, 0 to 19.1 (1.7) at 25%, 0 to 24.2 (2.8) at 10%, 0 to 29.9 (2.6) at 5% with SwiftScan. The contrast-to-noise ratios (CNR) were 95.3 at 75%, 88.3 at 50%, 69.2 at 25%, 45.7 at 10%, and 31.6 at 5% without SwiftScan, and 96.9, 91.7, 78.0, 71.6, and 62.0, respectively, using SwiftScan.

Conclusion: With the use of SwiftScan, a 50% reduction in acquisition time was considered acceptable for image quality with reproducible quantitative indices such as SUV and CNR.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11682469PMC
http://dx.doi.org/10.22038/aojnmb.2024.76919.1543DOI Listing

Publication Analysis

Top Keywords

image quality
12
swiftscan
10
semiconductor camera
8
75% 50%
8
considered acceptable
8
suv changes
8
median 75%
8
75%
5
50%
5
improving image
4

Similar Publications

Efficacy and cost-effectiveness of an ACT and compassion-based intervention for women with breast cancer: study protocol of two randomised controlled trials {1}.

Trials

January 2025

Center for Research in Neuropsychology and Cognitive and Behavioral Intervention, Faculty of Psychology and Education Sciences, University of Coimbra, Coimbra, Portugal.

Background: Breast cancer is the most diagnosed cancer in women worldwide and carries a considerable psychosocial burden. Interventions based on Acceptance and Commitment Therapy (ACT) and compassion-based approaches show promise in improving adjustment and quality of life in people with cancer. The Mind programme is an integrative ACT and compassion-based intervention tailored for women with breast cancer, which aims to prepare women for survivorship by promoting psychological flexibility and self-compassion.

View Article and Find Full Text PDF

Hyperspectral imaging (HSI) systems acquire images with spectral information over a wide range of wavelengths but are often affected by chromatic and other optical aberrations that degrade image quality. Deconvolution algorithms can improve the spatial resolution of HSI systems, yet retrieving the point spread function (PSF) is a crucial and challenging step. To address this challenge, we have developed a method for PSF estimation in HSI systems based on computed wavefronts.

View Article and Find Full Text PDF

We extend existing techniques by using generative adversarial network (GAN) models to reduce the appearance of cast shadows in radiographs across various age groups. We retrospectively collected 11,500 adult and paediatric wrist radiographs, evenly divided between those with and without casts. The test subset consisted of 750 radiographs with cast and 750 without cast.

View Article and Find Full Text PDF

Frequently, we perceive emotional information through multiple channels (e.g., face, voice, posture).

View Article and Find Full Text PDF

Thyroid surgery under nerve auto-fluorescence & artificial intelligence tissue identification software guidance.

Langenbecks Arch Surg

January 2025

Department of General Surgery, Sanatorio Otamendi & Miroli (Otamendi & Miroli Hospital), University of Buenos Aires, Buenos Aires, Argentina.

Thyroid cancer is a common malignancy that requires comprehensive clinical evaluation prior to adequate surgical management. Over the last three decades thyroid surgery has tripled and is considered one of the most commonly performed procedures in general surgery. These procedures are associated with potential postoperative complications with significant deterioration in the patient's quality of life.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!