Background: Achieving hemostasis of large bore venous access sites can be challenging and time consuming. Closure devices have proven to be superior in achieving hemostasis, reducing time to ambulation and improving patient comfort, compared to manual hemostasis techniques after femoral venous and arterial access. The closure of the jugular vein following large bore access has not been investigated in previous studies. In addition, scar formation of the neck after large bore access of the jugular vein has not been investigated. In this study, the safety and feasibility of the double Perclose ProGlide (PP), for achieving hemostasis of the internal jugular vein (IJV) following large bore access with 27 French Micra Transcatheter Pacemaker System (TPS) was examined. Also, the scar formation in the neck after IJV closure was examined during follow-up.

Methods: 136 consecutive patients from May 2018 until June 2024, in whom the IJV was closed with a double PP, following Micra TPS implantation were included. All patients were examined for hemostasis of the IJV and vascular complications, resulting in additional interventions. Time to ambulation, discharge and patient discomfort were also assessed. During follow-up the scar formation of the neck was examined.

Results: In all patients, the double PP was successful in achieving acute hemostasis of the IJV after large bore access. In all patients, 2 PP were deployed without device failure. One patient required additional manual pressure due to a minor hematoma. Ultrasound guided examination did not reveal any vascular complications. All patients were ambulated immediately. During follow-up, the scar in the neck was hardly visible.

Conclusions: Although the PP was designed as a closure device for femoral venous and arterial access, our data suggest that the PP can be used safely as a closure device for the IJV to achieve acute hemostasis, facilitate direct ambulation and improve patient comfort.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11683715PMC
http://dx.doi.org/10.31083/j.rcm2512440DOI Listing

Publication Analysis

Top Keywords

large bore
20
scar formation
16
bore access
16
achieving hemostasis
12
jugular vein
12
formation neck
12
time ambulation
8
patient comfort
8
femoral venous
8
venous arterial
8

Similar Publications

Transcatheter mass extraction of left-sided cardiac masses has gained popularity in recent years, with scarce data on effectiveness, safety, and types of devices used. Mostly, left-sided aspirations use mechanical and continuous-flow-mediated devices (the AngioVac [AngioDynamics] and Penumbra systems [Penumbra]). To our knowledge, the use of manual aspiration devices has not been reported yet.

View Article and Find Full Text PDF

We report the case of a previously independent 82-year-old female who experienced acute hemodynamic and respiratory deterioration requiring inotropic support due to a fat embolism during revision hip arthroplasty. Computed tomography pulmonary angiography demonstrated fat embolism, and transesophageal echocardiogram showed evidence of right ventricle strain and fat embolism in-transit in the right heart, as well as a moderate patent foramen ovale. Under transesophageal echocardiogram and intravascular ultrasound guidance, the Inari FlowTriever thrombectomy device was used successfully to retrieve the fat embolism with immediate hemodynamic improvement, no complications, and uneventful recovery.

View Article and Find Full Text PDF

Safety of accessing brachial veins for large-bore upper extremity venous thrombectomy using ClotTriever Thrombectomy System.

CVIR Endovasc

January 2025

Department of Radiology, Section of Vascular and Interventional Radiology, University of Washington, 1959 Northeast Pacific Street, Seattle, WA, 98195, USA.

Purpose: To evaluate access site adverse events following ClotTriever-mediated large-bore mechanical thrombectomy via small upper extremity deep veins (<ā€‰6-mm).

Materials And Methods: Twenty patients, including 24 upper extremity venous access sites, underwent ClotTriever-mediated large-bore thrombectomy of the upper extremity and thoracic central veins for symptomatic deep vein obstruction unresponsive to anticoagulation. Patients without follow-up venous duplex examinations (nā€‰=ā€‰3) were excluded.

View Article and Find Full Text PDF

Objective: In-bore MRI-guided biopsy allows direct visualization of suspicious lesions, biopsy needles, and trajectories, allowing accurate sampling when MRI-ultrasound fusion biopsy is not feasible. However, its use has been limited. Wide-bore, lower-field, and lower-cost scanners could help address these issues, but their feasibility for prostate biopsy is unknown.

View Article and Find Full Text PDF

Purpose: To describe the downsizing post-closure technique for access hemostasis during emergency endovascular repair (EVAR) in ruptured abdominal aortic aneurysms (RAAA).

Materials And Methods: A cohort of eight patients underwent emergency EVAR through 16 femoral access sites for infrarenal RAAA. The downsizing post-closure technique, which involves a reduction in the size of the large-bore access by advancing a 10F sheath, was consistently applied.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!