Steerable Delivery Sheath for Optimized LAA Closure: First Experience and Procedural Outcomes.

Catheter Cardiovasc Interv

Department of Cardiology, Angiology and Intensive Care Medicine, University Heart Center Lübeck, Lübeck, Germany.

Published: December 2024

Background: The new Amplatzer Steerable Delivery Sheath is a delivery system designed to improve ease-of-use and procedural results of left atrial appendage closure (LAAC). We aimed to compare procedural results after switching our LAAC program at a tertiary care center with the Amulet device to the Steerable Delivery Sheath, with a control group of LAAC employing the standard sheath.

Methods: The first n = 32 consecutively treated patients at our site using the Amulet device with the Steerable Delivery Sheath were included in this retrospective analysis. As a control-group, n = 39 consecutive patients treated with the Amulet device before the switch to the new sheath were used.

Results: LAAC was successful in all patients in both groups (100%). Contrast use and fluoroscopy time were numerically higher in the steerable sheath group (steerable sheath vs. control group: contrast use 70 ± 23 vs. 55 ± 50 mL, p = 0.09, fluoroscopy time 12.7 ± 4.6 vs. 10.2 ± 6.6 min, p = 0.3). Fluoroscopy time and dose decreased after 3 months while contrast use remained unchanged. Complete sealing rate was high in both groups (steerable sheath vs. control group 97% vs. 95%, p > 0.9) and periprocedural complication rate was low, without any periprocedural stroke, vascular complications, or death in both cohorts.

Conclusions: LAAC with the Amplatzer Amulet steerable delivery sheath is feasible and safe. Fluoroscopy time and dose suggest a learning curve with the new sheath.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.31390DOI Listing

Publication Analysis

Top Keywords

steerable delivery
20
delivery sheath
20
fluoroscopy time
16
amulet device
12
sheath control
12
control group
12
steerable sheath
12
sheath
10
steerable
8
device steerable
8

Similar Publications

Background: The new Amplatzer Steerable Delivery Sheath is a delivery system designed to improve ease-of-use and procedural results of left atrial appendage closure (LAAC). We aimed to compare procedural results after switching our LAAC program at a tertiary care center with the Amulet device to the Steerable Delivery Sheath, with a control group of LAAC employing the standard sheath.

Methods: The first n = 32 consecutively treated patients at our site using the Amulet device with the Steerable Delivery Sheath were included in this retrospective analysis.

View Article and Find Full Text PDF
Article Synopsis
  • - The study examines the safety and efficacy of a new 8F variable loop multielectrode catheter for pulsed field ablation (PFA) in patients with atrial fibrillation, which simplifies catheter use compared to earlier complex designs.
  • - Out of 30 patients, all achieved complete pulmonary vein and posterior wall isolation, with an average procedural time of about 113 minutes and minimal complications, including a few transient cerebral lesions.
  • - The research highlights that this new catheter system provides effective PFA while maintaining a safety profile similar to existing options, showing promise for future use in treating atrial fibrillation.
View Article and Find Full Text PDF

Objectives: Transcranial low-intensity focused ultrasound (LIFU) offers unique opportunities for precisely neuromodulating small and/or deep targets within the human brain, which may be useful for treating psychiatric and neurological disorders. This article presents a novel ultrasound system that delivers focused ultrasound through the forehead to anterior brain targets and evaluates its safety and usability in a volunteer study.

Methods: The ultrasound system and workflow are described, including neuronavigation, LIFU planning, and ultrasound delivery components.

View Article and Find Full Text PDF

Background: Enlarged left atrium (LA) is frequently identified in patients who undergo left atrial appendage closure (LAAC) and negatively affects the device's final position and intervention results. Steerable delivery sheath (SDS) could represent an option to overcome these difficulties. We aimed to assess the feasibility of SDS-assisted procedures and compare their efficacy to the standard sheath strategy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!