Objective: To explore weight-bearing stability of Pilon fracture fixed by external fixator.
Methods: Six ankle bone models (right side) and 4 pairs (8 ankle cadaver specimens) were selected. Pilon fracture model was prepared by using the preset osteotomy line based on Ruedi Allgower Pilon fracture type. Pilon fracture model was built by using a minimally invasive osteotomy. After ankle bone model and cadaver specimen model were fixed with external fixator, axial load was carried out on mechanical loading machine. After ankle bone model and cadaver specimen model were fixed with external fixator, axial load was carried out on mechanical loading machine. Axial loads of 150, 300 and 450 N were applied to ankle bone model, and displacements of fibula fracture blocks, lateral tibia fracture blocks and medial tibia fracture blocks in three-dimensional space (X, Y and Z axes) were recorded by dynamic capture instrument. Axial loads of 300, 600 and 900 N were applied to ankle cadaver model fixed by external fixator. X-ray films of Pilon fracture cadaver model fixed by external fixator under different loading conditions were taken. The anterior tibial angle, tibial malleolar point angle, talus shift value, talus tilt angle, lateral malleolar shift value, lateral malleolar shift value, medial malleolar separation shift value and articular surface step displacement value were measured under different loads by digimizer software.
Results: After 150, 300 and 450 N axial loads were applied to Pilon fracture models fixed by external fixator, no loosening or fracture of external fixator was observed, and no loosening, fracture or irreversible plastic deformation of Kirschner needle were observed. The displacement values of fibular fracture pieces on X-axis(around) were 0.032 (-0.022, 0.269), 0.061 (-0.002, 0.427), 0.212(-0.016, 1.223) mm, and the displacement values on Y-axis(above and below) were 0.002(-0.031, 0.103), 0.051(-1.133, 0.376), 0.128 (-1.394, 0.516) mm, and displacement values on Z-axis (front and rear) were -0.003 (-0.130, 0.171), 0.137 (-0.076, 0.433), 0.030(-0.487, 0.478) mm;the displacement values of lateral tibial fractures on X-axis were 0.000(-0.108, 0.027), 0.083(-0.364, 0.050), -0.121(-0.289, 0.165) mm, and displacement values on Y-axis were -0.009(-0.200, 0.025), -0.179(-0.710, 0.084), -0.257(-0.799, 0.027) mm, and displacement values on Z-axis were 0.112(-0.024, 0.256), 0.157(-0.068, 0.293), -0.210(-0.035, 0.430) mm;the displacement values of medial tibial fracture block on X-axis were -0.010(-0.060, 0.013), -0.165(-0.289, 0.056), -0.181(-0.395, 0.013) mm, and the displacement values on Y-axis were -0.036(-0.156, 0.007), -0.104(-0.269, 0.178), -0.245(-0.380, -0.011) mm, and displacement values on Z-axis were -0.005(-0.372, 0.189), -0.012 (-1.774, 0.380), 0.200 (-1.963, -0.540) mm. After 300, 600 and 900 N axial loads were applied to Pilon fracture cadaverous models fixed with external fixators, there were no significant difference in anterior tibial angles, angles of malleolar points of tibia, oblique angles of talus, fracture steps, shift values of talus, lateral shift values of lateral malleolus, lateral shift values of medial malleolus, lateral shift values of medial malleolus between under different loading conditions and those without loading (>0.05). No loosening or fracture of external fixator as a whole, loosening, fracture or irreversible deformation of Kirschner needle at the local fixed fracture end occurred.
Conclusion: The early weight-bearing external fixator could maintain stability of fracture end and ankle joint, and the maximum weight is not more than 300 N. In clinical practical application, material characteristics of the implant and type of fracture should be selected.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.12200/j.issn.1003-0034.20230702 | DOI Listing |
Am J Orthod Dentofacial Orthop
January 2025
Department of Orthodontics, Faculty of Dentistry, Hacettepe University, Ankara, Turkey. Electronic address:
Introduction: The objective of this study was to evaluate the effects of the miniplate application sites in the maxilla and the applied force vector changes during skeletally supported facemask application in adolescent patients with unilateral cleft lip and palate (UCLP) using finite element model (FEM) analysis.
Methods: A FEM was obtained from a cone-beam computed tomography image of a 12-year-old female patient with UCLP. Miniplates were placed on 3 different sites of the maxilla; 500 g of advancement force was applied bilaterally, parallel (0°), and downward (-30°) to the occlusal plane.
Materials (Basel)
January 2025
Faculty of Furniture Design and Wood Engineering, Transilvania University of Brasov, B-dul Eroilor, nr. 29, 500036 Brasov, Romania.
The present paper investigates the possibility of replacing the traditional L-type corner joint used in chair construction with a 3D printed connector, manufactured using the Fused Filament Fabrication (FFF) method and black PLA as filament. The connector was designed to assemble the legs with seat rails and stretchers, and it was tested under diagonal tensile and compression loads. Its performance was compared to that of the traditional mortise-and-tenon joint.
View Article and Find Full Text PDFMaterials (Basel)
January 2025
Materials Research Centre for Energy and Clean Technology, School of Materials Science and Engineering, Andong National University, Andong 36729, Republic of Korea.
The accurate measurement of the elastic modulus of thin metal films on flexible substrates is critical for understanding the mechanical reliability of flexible electronics. However, conventional methods, such as the Oliver-Pharr model, often underestimate the modulus due to substrate effects, particularly with low-modulus substrates like polyimide (PI). In this study, we propose an improved weighting model that replaces the empirical weighting factor with a variable X to better account for substrate contributions.
View Article and Find Full Text PDFMaterials (Basel)
January 2025
Faculty of Mechanical Engineering and Aeronautics, Rzeszow University of Technology, 35-959 Rzeszow, Poland.
This article presents an investigation of the use of machine learning methodologies for the prediction of surface roughness in milling operations, using sensor data as the primary source of information. The sensors, which included current transformers, a microphone, and displacement sensors, captured comprehensive machining signals at a frequency of 10 kHz. The signals were subjected to preprocessing using the Savitzky-Golay filter, with the objective of isolating relevant moments of active material machining and reducing noise.
View Article and Find Full Text PDFJ Orthop Surg Res
January 2025
Hebei Orthopaedic Research Institute, Hebei Medical University Third Hospital, No.139 Ziqiang Road, Shijiazhuang, 050051, P.R. China.
Objective: The postoperative recovery of patients with lumbar disc herniation (LDH) requires further study. This study aimed to establish and validate a predictive model for functional recovery in patients with LDH and explore associated risk factors.
Method: Patients with LDH undergoing PLIF admitted from January 1, 2018 to December 31, 2022 were included, and patient data were prospectively collected through follow-up.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!