Background: The cementation technique is crucial for achieving adequate fixation and optimal survivorship in total knee arthroplasty (TKA). The thickness of the cement at the tibial bone-implant surface may be related to aseptic tibial loosening. However, to date, no studies have demonstrated a direct association between cement thickness and rates of aseptic tibial loosening.

Methods: We performed a retrospective review to identify 28,327 primary cemented TKAs with at least 2 years of follow-up at an academic health system from 2013 to 2021. A total of 115 cases underwent revision surgery for aseptic tibial loosening. Cases where the implant was recalled specifically for loosening (n = 23) were excluded. The remaining 92 aseptic tibial loosening cases were 2:1 propensity score matched and implant matched to control patients who did not have tibial loosening. There were two independent reviewers who then measured the thickness of the cement interface in 10 locations along the bone-implant interface from initial postoperative radiographs. The averages of the reviewers' measurements were calculated and then compared using independent t-tests.

Results: Aseptic tibial loosening cases involving implant A tibial baseplate (n = 75) had significantly thinner cement interfaces than matched controls at all the 10 locations measured. Aseptic loosening cases involving implant B (n = 17) also displayed a thinner cement interface than matched controls in all locations, but this result was only statistically significant at the medial baseplate, medial keel, lateral keel, anterior keel, and posterior baseplate.

Conclusions: In two widely used TKA systems, tibial aseptic loosening was associated with significantly thinner cement interfaces when compared to propensity-matched controls in two different implant types. Further prospective studies are needed to identify the optimal keel preparation and design as well as minimal cement interface thickness to avoid implant loosening.

Level Of Evidence: III.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arth.2024.12.023DOI Listing

Publication Analysis

Top Keywords

aseptic tibial
24
tibial loosening
24
loosening cases
16
thickness cement
12
cement interface
12
thinner cement
12
loosening
9
tibial
9
aseptic
8
loosening associated
8

Similar Publications

Background: Total knee arthroplasty (TKA) is one of the most common orthopaedic procedures and the number of patients which undergo TKA will continue to rise in the coming years. Consecutively, the number of necessary revision surgeries will increase. One of the main reasons for revision surgery is aseptic loosening because of a so-called stress-shielding effect.

View Article and Find Full Text PDF

Introduction: Aseptic recalcitrant nonunion (ARNU) of the femur and tibia is an entity in which the absence of bony union, misalignment, and limb length discrepancies (LLD) coexist. Currently, the management of these cases lacks consensus. This study aimed to describe the bone union rate and deformity correction outcomes in patients with ARNU of the femur or tibia treated with the Induced Membrane Technique (IMT).

View Article and Find Full Text PDF

Background: Coronal lower extremity malalignment and improper implant position have been described as risk factors for aseptic tibial loosening following primary total knee arthroplasty (TKA). However, several prior studies have shown no association between alignment and implant loosening. Meanwhile, there is increasing interest in kinematic alignment.

View Article and Find Full Text PDF

Purpose: >The aim of this multicenter study was to analyze the potential impact of patient demographics and cementation technique towards the development of radiolucent lines (RLLs) in primary total knee arthroplasty (TKA). It was hypothesized that cementation techniques, including higher cement volume, double-layer cementation technique and hardening in full extension, reduce RLL incidence by improving stability, whereas demographic factors such as age, BMI and smoking may increase RLL risk by affecting bone quality and mechanical loading.

Methods: Altogether, 776 patients (median age: 70.

View Article and Find Full Text PDF

Background And Purpose:  Early migration of the uncemented cruciate-sacrificing rotating platform ATTUNE and Low Contact Stress (LCS) tibial components was classified as at-risk for aseptic loosening rates exceeding 6.5% at 15 years based on recent fixation-specific migration thresholds. In this secondary report of a randomized controlled trial (RCT) we aimed to evaluate whether the 5-year migration, inducible displacement, and the clinical outcome of the ATTUNE components were comparable to those of the LCS.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?