Background: The study assesses the feasibility of the DuoCor BiVAS, a novel biventricular assist system integrating magnetic levitation technology.

Methods: In an acute large animal model involving five sheep, each received the DuoCor BiVAS without cardiopulmonary bypass. Hemodynamic and device parameters were monitored continuously for 1-h post-implantation.

Results: Intraoperative implantation was uneventful, demonstrating successful biventricular support with mean blood flows of 5.4 LPM (left) and 5.5 LPM (right). Analysis showed proportional flow rates and relationships between pump speed, flow, and power consumption. No adverse events like thrombus formation, bleeding, stroke, or device failure occurred.

Conclusion: This research underscores the DuoCor BiVAS's potential for severe biventricular heart failure treatment, providing insights into its feasibility and functionality in acute animal models. The findings suggest promising clinical applications, particularly with the system's single driveline design potentially enhancing patient mobilization and quality of life. Further investigations are needed to advance this technology for broader clinical use.

Download full-text PDF

Source
http://dx.doi.org/10.1111/aor.14933DOI Listing

Publication Analysis

Top Keywords

duocor bivas
8
coupled continuous-flow
4
continuous-flow ventricular
4
ventricular assist
4
assist devices
4
devices novel
4
novel bivad
4
bivad driveline-acute
4
driveline-acute animal
4
animal study
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!