Characterization of Feline Basophils on the Sysmex XN-1000V and Evaluation of a New WDF Gating Profile.

Animals (Basel)

Servei d'Hematologia Clínica Veterinària, Departament de Medicina i Cirurgia Animals, Facultat de Veterinària, Universitat Autònoma de Barcelona, 08193 Bellaterra, Spain.

Published: November 2024

The Sysmex XN-1000V provides a percentage and concentration of basophils from the WNR scattergram, as for human samples, but this method has been shown to be irrelevant in cats. This study aimed to characterize the feline basophil distribution on the WDF channel and to preliminarily evaluate the performance of a new basophil gate on the WDF channel. Cases of feline basophilia were retrospectively retrieved and the scattergram from the WDF and WNR channels were evaluated for any consistent pattern. A new gating setting for the WDF channel was created to include the suspected basophil region. This new gating was applied retrospectively to identified basophilia cases and prospectively to randomly selected feline cases. Manual, WNR, and new WDF methods for basophil identification were compared. Nine cases of feline basophilia were identified. A characteristic WDF scattergram was identified in seven of the nine cases. A new gate was created on the WDF channel and applied to these and 34 additional cases. The comparison study showed that the new method of basophil quantification using the WDF scattergram correlated better with the manual method than the Sysmex XN-1000V method using the WNR scattergram. Basophil concentration in feline peripheral blood can be determined using a new gate on the WDF channel of the Sysmex XN-1000V, which provides better performance than the WNR channel and is comparable to the manual method.

Download full-text PDF

Source
http://dx.doi.org/10.3390/ani14233362DOI Listing
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11639760PMC

Publication Analysis

Top Keywords

wdf channel
20
sysmex xn-1000v
16
wdf
10
wnr scattergram
8
gate wdf
8
cases feline
8
feline basophilia
8
wdf scattergram
8
manual method
8
basophil
6

Similar Publications

Characterization of Feline Basophils on the Sysmex XN-1000V and Evaluation of a New WDF Gating Profile.

Animals (Basel)

November 2024

Servei d'Hematologia Clínica Veterinària, Departament de Medicina i Cirurgia Animals, Facultat de Veterinària, Universitat Autònoma de Barcelona, 08193 Bellaterra, Spain.

The Sysmex XN-1000V provides a percentage and concentration of basophils from the WNR scattergram, as for human samples, but this method has been shown to be irrelevant in cats. This study aimed to characterize the feline basophil distribution on the WDF channel and to preliminarily evaluate the performance of a new basophil gate on the WDF channel. Cases of feline basophilia were retrospectively retrieved and the scattergram from the WDF and WNR channels were evaluated for any consistent pattern.

View Article and Find Full Text PDF

Background: The new XR-Series haematology analyser from Sysmex provides increased throughput and automation, along with a new reagent in WDF channel for optimised WBC differential.

Methods: An analytical performance study for the XR analyser was conducted to evaluate the WDF channel parameters in comparison to the instrument specifications. Additionally, 7460 samples were measured on XR and XN analysers to compare selected parameters and flags, and 930 randomly selected samples were further evaluated with microscopy.

View Article and Find Full Text PDF

Introduction: This study investigates the potential contribution of structural and dispersion parameters, as well as alarms provided by Sysmex XN9000 haematology analyzer. The objective was to assess the need for a microscopic examination in the context of lymphocytosis. It also aims to contribute in differentiating rapidly lymphoproliferative disorders such as chronic lymphocytic leukaemia (CLL), non-chronic lymphocytic leukaemia (NON-CLL) and non-infectious reactive lymphocytosis (REAC).

View Article and Find Full Text PDF

Introduction: This study evaluated the feasibility of the Sysmex XN-3000 automated hematology analyzer for the assessment of total nucleated cells (TNC) and bone marrow (BM) cell density in routine bone marrow aspiration (BMA) samples.

Methods: A total of 54 BMA samples from 39 hematological patients were evaluated. The number of megakaryocytes was calculated by a specific gating algorithm using the body fluid mode of the WBC differential (WDF) channel.

View Article and Find Full Text PDF

Background: Hematology analyzers provide quick and accurate results in most situations. However, spurious results related to the parameters from the complete blood count (CBC) may be observed in several instances. False increases in WBC count can occur for many reasons, including erythroblasts, insufficiently lysed red blood cells (RBC), platelet aggregates, lipids, and cryoglobulins.