. To investigate the effect of the position and orientation of the detector and its influence on the determination of output factors (OF) for small fields for a linear accelerator (MR-linac) integrated with 1.5 T magnetic resonance following the TRS-483 formalism.. OF were measured for small fields in the central axis following the recommendations of the manufacturer and at the dose maximum following the TRS-483 formalism. OF were determined using a microDiamond (MD), a Semiflex (SF) 31021 ionization chamber, Gafchromic EBT3 film and were calculated in Monaco treatment planning system (TPS). Additionally, the orientation response of SF was evaluated, placing it in parallel and perpendicular direction to the radiation beam. The values were compared taking film measurements as reference. The corrected factors,ΩQclinical,msrfclinical,msr, required the use of output correction factorkQclinical,msrfclinical,msrtaken from previous reports. Finally, there are proposed experimentalkQclinical,msrfclinical,msrfor SF and MD, following the measured values in this work.. In fields smaller than 4 cm, the positioning of the SF and MD in the central axis or at the point of dose maximum affects the reading significantly with differences of up to 6% and 4%, respectively. For the data calculated in the TPS, the maximum difference of the OF between MD and TPS for fields greater than 2 cm was 0.6% and below this field size the TPS underestimates the OF up to 10.6%. The orientation (parallel or perpendicular) of the SF regarding the radiation beam has a considerable impact on the OF for fields smaller than 3 cm, showing a variation up to 10% for the field of 0.5 cm.. This study provides valuable information on the challenges and limitations of measuring output factors in small fields. The outcomes have important implications for the practice of radiosurgery, underscoring the need for accuracy in detector placement and orientation, as well as the importance of using more advanced technologies and more robust measurement methods.

Download full-text PDF

Source
http://dx.doi.org/10.1088/2057-1976/ad9f67DOI Listing

Publication Analysis

Top Keywords

small fields
16
output factors
12
position orientation
8
orientation detector
8
factors small
8
trs-483 formalism
8
central axis
8
dose maximum
8
parallel perpendicular
8
radiation beam
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!