  • The study focuses on measuring beam alignment in medical linear accelerators, emphasizing the importance of analyzing focal spot positions to ensure accurate radiation delivery.
  • The newly developed Quality Control Tool (QCT) device facilitates simultaneous tests of light/radiation field coincidence and isocenter testing, enabling effective analysis of focal spots with specific software.
  • Results show that focal spot position varies with gantry angle, showcasing the QCT's capacity to accurately measure displacements and assess beam alignment without interference from the jaw collimator.

Article Abstract

Purpose: Measurement of beam alignment, including an isocenter, is a fundamental part of commissioning and quality assurance for a medical linear accelerator (linac). An alignment shift is caused by the focal spot displacement from the beam axis. In this study, we present a procedure for the analysis of the focal spot positions in the radiation alignment using a newly developed quality control tool (QCT) device and analysis software.

Methods: The device has the function to perform the light/radiation field coincidence test, and the isocenter test at the same time. First, the light/radiation field was evaluated by using cross-slits on the surface of the QCT device. Second, the focal spot positions of beam alignment were measured using QCT, and the acquired images were collectively analyzed using the developed analysis software. Finally, measurements of focal spot positions with different energies were made using the accelerators at the 2 facilities.

Results: The focal spot position was changed with a gantry angle. The focal spot displacement was 0.15-1.37 mm. With the validation of 2 accelerators at 2 facilities, the displacement of the focal spot position could be measured using the QCT phantom.

Conclusion: The focal spot position measurement using the QCT test captured the change in the beam alignment between the beam axes without being affected by the jaw collimator. This study suggested that focal spot position measurement using a QCT device is useful for evaluating beam alignment at linac.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.25-1472DOI Listing

Publication Analysis

Top Keywords

focal spot
36
beam alignment
16
spot position
16
spot positions
12
qct device
12
focal
9
spot
9
medical linear
8
spot displacement
8
light/radiation field
8

Similar Publications

Nanoparticle-mediated light-driven LAMP combined with test strips for sensitive and rapid visual detection of antibiotic resistance genes.

J Hazard Mater

December 2024

State Key Laboratory of Food Science and Resources, Jiangnan University, Wuxi 214122, China; International Joint Laboratory on Food Safety, Institute of Analytical Food Safety, School of Food Science and Technology, Jiangnan University, Wuxi, Jiangsu 214122, China. Electronic address:

Antibiotic resistance genes (ARGs) are markers of drug-resistant pathogens, monitoring them contributes to prevent resistance to drugs. The detection methods for ARGs including PCR and isothermal amplification are sensitive and selective. However, it may take several hours or cannot be used on spot.

View Article and Find Full Text PDF

To balance the convergence speed and solution diversity and enhance optimization performance when addressing large-scale optimization problems, this research study presents an improved ant colony optimization (ICMPACO) technique. Its foundations include the co-evolution mechanism, the multi-population strategy, the pheromone diffusion mechanism, and the pheromone updating method. The suggested ICMPACO approach separates the ant population into elite and common categories and breaks the optimization problem into several sub-problems to boost the convergence rate and prevent slipping into the local optimum value.

View Article and Find Full Text PDF

Background: Highly pathogenic avian influenza (HPAI) (H5N1) has been endemic in Egypt for almost two decades, profoundly impacting both the poultry industry and public health. Egypt stands as a prominent epicenter for HPAI H5N1 outbreaks in Africa, marked by the highest number of positive human cases. Despite continuous governmental efforts, prior research underscored the inadequacy of strategies in controlling the virus spread.

View Article and Find Full Text PDF

Background: Online grocery shopping is a growing source of food purchases in many countries. We investigated the effect of nudging consumers towards purchases of lower sodium products using a web browser extension.

Methods: This trial was conducted among individuals with hypertension who shopped for their groceries online in Australia.

View Article and Find Full Text PDF

Background: This study aimed to evaluate the diagnostic performance of ESAT6-CFP10 (EC) skin test in healthy population and determine the factors influencing the booster effect.

Methods: We conducted a randomized, double-blind, parallel controlled trial in healthy population. The experiment was divided into two stages.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!