Differences in magnetic resonance (MR) imaging signal-to-noise (S/N) performance between saddle-shaped and solenoidal coils have been postulated. Each coil shape is tied to a particular magnetic field configuration, so that they are not typically interchangeable except in special situations. The solenoidal coil is predicted to have a two- to three-fold advantage over the saddle-shaped coil. Simple basic arguments raise a dispute with this assertion. Experiments show that both coils produce essentially equivalent S/N levels.