Purpose: B field inhomogeneity within the brain-feeding arteries is a major issue for pseudo-continuous arterial spin labeling (PCASL) at 7 T because it reduces the labeling efficiency and leads to a loss of perfusion signal. This study aimed to develop a vessel-specific dynamic B field shimming method for 7 T PCASL to improve the labeling efficiency by correcting off-resonance within the arteries in the labeling region.
Methods: We implemented a PCASL sequence with dynamic B shimming at 7 T that compensates for B field offsets in the brain-feeding arteries by updating linear shimming terms and adding a phase increment to the PCASL RF pulses. Rapidly acquired vessel-specific B field maps were used to calculate dynamic B shimming parameters. We evaluated both 2D and 3D variants of our method, comparing their performance against the established global frequency offset and optimal encoding scheme-based corrections. Cerebral blood flow (CBF) maps were quantified before and after corrections, and CBF values from different methods were compared across the whole brain, white matter, and gray matter regions.
Results: All off-resonance correction methods significantly recovered perfusion signals across the brain. The proposed vessel-specific dynamic B shimming method improved the labeling efficiency while maintaining optimal static shimming in the imaging region. Perfusion-weighted images demonstrated the superiority of the 3D dynamic B shimming method compared to global or 2D-based correction approaches. CBF analysis revealed that 3D dynamic B shimming significantly increased CBF values relative to the other methods.
Conclusion: Our proposed dynamic B shimming method offers a significant advancement in PCASL robustness and effectiveness, enabling full utilization of 7 T ASL high sensitivity and spatial resolution.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mrm.30387 | DOI Listing |
Dentomaxillofac Radiol
January 2025
Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing 100730, China.
Objectives: To compare integrated slice-specific dynamic shimming (iShim) and simultaneous multi-slice (SMS) readout-segmented echo-planar imaging (RESOLVE) for diffusion-weighted imaging (DWI) of malignant head and neck tumours.
Methods: In this prospective study, 45 patients with malignant head and neck lesions underwent iShim- and SMS-RESOLVE imaging with two b-values (0, 800 s/mm2) at 3 T. Subjective image quality scores (lesion distortion, signal loss, fat saturation, and artefacts), quantitative lesion distortion, quantitative image quality [signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and SNR efficiency], ADC values, and total acquisition times of iShim- and SMS-RESOLVE imaging were evaluated and compared.
Magn Reson Med
December 2024
Wellcome Centre for Integrative Neuroimaging, FMRIB Division, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
Purpose: B field inhomogeneity within the brain-feeding arteries is a major issue for pseudo-continuous arterial spin labeling (PCASL) at 7 T because it reduces the labeling efficiency and leads to a loss of perfusion signal. This study aimed to develop a vessel-specific dynamic B field shimming method for 7 T PCASL to improve the labeling efficiency by correcting off-resonance within the arteries in the labeling region.
Methods: We implemented a PCASL sequence with dynamic B shimming at 7 T that compensates for B field offsets in the brain-feeding arteries by updating linear shimming terms and adding a phase increment to the PCASL RF pulses.
Med Phys
November 2024
School of Information Technology and Electrical Engineering, The University of Queensland, Brisbane, Australia.
Background: In magnetic resonance imaging (MRI), maintaining a highly uniform main magnetic field (B) is essential for producing detailed images of human anatomy. Passive shimming (PS) is a technique used to enhance B uniformity by strategically arranging shimming iron pieces inside the magnet bore. Traditionally, PS optimization has been implemented using linear programming (LP), posing challenges in balancing field quality with the quantity of iron used for shimming.
View Article and Find Full Text PDFPsychoradiology
August 2024
Department of Radiology, Weill Medical College of Cornell University, New York, NY 10065, United States.
High magnetic field homogeneity is critical for magnetic resonance imaging (MRI), functional MRI, and magnetic resonance spectroscopy (MRS) applications. B inhomogeneity during MR scans is a long-standing problem resulting from magnet imperfections and site conditions, with the main issue being the inhomogeneity across the human body caused by differences in magnetic susceptibilities between tissues, resulting in signal loss, image distortion, and poor spectral resolution. Through a combination of passive and active shim techniques, as well as technological advances employing multi-coil techniques, optimal coil design, motion tracking, and real-time modifications, improved field homogeneity and image quality have been achieved in MRI/MRS.
View Article and Find Full Text PDFCancer Imaging
August 2024
MR Scientific Marketing, Siemens Healthineers, Shanghai, 201318, China.
Background: The aim of this research is to prospectively investigate the diagnostic performance of intravoxel incoherent motion (IVIM) using the integrated slice-specific dynamic shimming (iShim) technique in staging primary esophageal squamous cell carcinoma (ESCC) and predicting presence of lymph node metastases from ESCC.
Methods: Sixty-three patients with ESCC were prospectively enrolled from April 2016 to April 2019. MR and IVIM using iShim technique (b = 0, 25, 50, 75, 100, 200, 400, 600, 800 s/mm) were performed on 3.