  • Robotics in neurosurgery is increasingly used, particularly with the incorporation of intraoperative imaging for precise patient registration in a new lightweight, table-mounted robotic system designed for cranial biopsies and sEEG procedures.
  • The study evaluated 26 patients using the Cirq system for stereotactic biopsy or sEEG from January 2023 to August 2024, comparing results with data from the frameless VarioGuide system, specifically looking at entry errors, target errors, and vector deviation of electrodes.
  • Results indicated that Cirq-assisted procedures had a mean surgical time of about 117 minutes for biopsies and 269 minutes for sEEG, revealing differences in efficiency compared to VarioGuide, where the comparable

Article Abstract

Objective: Robotics in neurosurgery is becoming increasingly prevalent. The integration of intraoperative imaging for patient registration into workflows of newer robotic systems enhances precision and has further driven their widespread adoption. In this study, the authors report on a lightweight, table-mounted robotic system integrating robotic cone-beam CT (CB-CT) for automated patient registration in cranial biopsies and stereotactic electroencephalography (sEEG).

Methods: This prospective cohort study included patients who underwent stereotactic biopsy or sEEG with the Cirq system from January 2023 to August 2024. For patient-to-image registration, an external registration matrix was secured near the patient's head before conducting CB-CT with robotic Artis Pheno. CT was then fused with preoperative planning MRI and used as the navigation dataset. Demographic and clinical data were evaluated, and entry and target errors, as well as vector deviation of sEEG electrodes, were assessed and compared with those of patients who underwent biopsies and sEEG with the frameless VarioGuide system.

Results: In 26 Cirq-assisted surgical procedures, robotic CB-CT was used for image registration in 20 cases. Of these, 15 were biopsies (mean ± SD 7 ± 1 specimens) and 5 were sEEG with 31 depth electrodes, compared to 29 VarioGuide biopsies and 3 VarioGuide sEEG cases with 25 electrodes. The mean age was 56 ± 19 years, with a male/female ratio of 1.9:1. Lesion size averaged 19 ± 17 cm3 on T1-weighted imaging and 61 ± 53 cm3 on T2-weighted imaging for Cirq and 14 ± 14 cm3 and 68 ± 47 cm3 for VarioGuide. The mean surgical times were 117 ± 34 minutes for biopsy and 269 ± 54 minutes for sEEG in the Cirq group, with skin-to-skin times of 40 ± 23 minutes for biopsy and 208 ± 74 minutes for sEEG; in comparison, surgical times of 78 ± 21 minutes for biopsy and 218 ± 33 minutes for sEEG were reported with VarioGuide, with skin-to-skin times of 34 ± 13 and 158 ± 27 minutes. No complications occurred. The mean dosage area product was 983 ± 351 µGym2 for biopsies and 1772 ± 968 µGym2 for sEEG. Cirq-assisted sEEG electrodes had mean entry and target errors of 1.4 ± 1.2 mm and 2.6 ± 1.6 mm, compared to 5.3 ± 3.3 mm and 6.5 ± 2.8 mm with VarioGuide. Mean vector deviation was 1.6 ± 0.9 mm with Cirq versus 4.9 ± 2.9 mm with VarioGuide.

Conclusions: The integration of a lightweight, table-mounted robotic alignment tool with intraoperative CB-CT for automated patient-to-image registration enables high precision and a seamless workflow. This combination is safe, has a manageable learning curve, and holds potential to replace traditional frame-based and frameless procedures. Its efficiency and accuracy are likely to contribute to the increasing adoption of robotics in neurosurgery.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2024.9.FOCUS24525DOI Listing

Publication Analysis

Top Keywords

lightweight table-mounted
12
table-mounted robotic
12
patient-to-image registration
12
minutes biopsy
12
minutes seeg
12
seeg
10
integration lightweight
8
robotic
8
robotic alignment
8
alignment tool
8

Similar Publications

Article Synopsis
  • Robotics in neurosurgery is increasingly used, particularly with the incorporation of intraoperative imaging for precise patient registration in a new lightweight, table-mounted robotic system designed for cranial biopsies and sEEG procedures.
  • The study evaluated 26 patients using the Cirq system for stereotactic biopsy or sEEG from January 2023 to August 2024, comparing results with data from the frameless VarioGuide system, specifically looking at entry errors, target errors, and vector deviation of electrodes.
  • Results indicated that Cirq-assisted procedures had a mean surgical time of about 117 minutes for biopsies and 269 minutes for sEEG, revealing differences in efficiency compared to VarioGuide, where the comparable
View Article and Find Full Text PDF

TITAN CSR: a new self-retaining retractor for abdominal surgery.

Trauma Surg Acute Care Open

November 2023

Trauma and Emergency General Surgery, The University of Texas Health Science Center at San Antonio, San Antonio, Texas, USA.

Background: Self-retaining retractors provide hands-free exposure and allow surgeons to use their hands to perform surgery rather than manually retract tissues. Non-table-mounted retractors offer rapid assembly, but they have been traditionally limited in their ability to provide adequate exposure of the abdomen. Table-mounted retractors provide excellent exposure, but their cumbersome and time-intensive assembly often precludes their use in emergency settings.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!