A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Research on a novel flexible bionic artificial anal sphincter. | LitMetric

Research on a novel flexible bionic artificial anal sphincter.

Artif Organs

Institute of Rehabilitation Engineering and Technology, School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China.

Published: November 2024

Background: Artificial anal sphincter is a novel method for the treatment of fecal incontinence. Aiming at the problems of complex mechanical structure, high mechanical failure rate, large volume and poor biocompatibility of existing artificial anal sphincters, this article proposes a novel flexible artificial anal sphincter with bionic structure.

Methods: A novel flexible artificial anal sphincter was proposed with bionic structure by analogy with the puborectalis, flexible transmission by flexible tendons, good biocompatibility by using thermoplastic urethane and medical silicone, small size and lightweight. The interaction between the rectum and the artificial anal sphincter is simulated by the finite element analysis, and the changes of the anorectal angle and the stress during the traction process of the rectum are analyzed.

Results: The results suggest that the formed anorectal angle of <90° may effectively control defecation, so the effectiveness of the artificial anal sphincter is preliminarily verified. When the traction angle was 71° and the intestinal tube was basically closed, the maximum equivalent stress was 148.45 kPa, the maximum and minimum principal stress were 88.37 and 82.74 kPa, which were all below the safety pressure that initially verified the safety of the artificial anal sphincter.

Conclusion: The novel flexible artificial anal sphincter may effectively reduce the stress between the device and the rectum and improve the biomechanical compatibility.

Download full-text PDF

Source
http://dx.doi.org/10.1111/aor.14917DOI Listing

Publication Analysis

Top Keywords

artificial anal
24
anal sphincter
20
novel flexible
12
flexible artificial
8
anorectal angle
8
artificial
6
anal
6
sphincter
5
novel
4
flexible bionic
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!