In 2020, 20% of patients with COVID-19 developed severe complications, including life-threatening pneumonia with systemic inflammatory response syndrome (SIRS). We developed a preliminary SIRS monitor that does not require blood sampling, is noninvasive, and can collect data 24 h per day. The proposed monitor comprises a piezoelectric respiratory sensor located beneath the patient's mattress and a fingertip pulse sensor that determines ultra-high accuracy respiratory rate (mode of a 40-min frequency distribution of respiratory rates (M40FD-RR)). We assessed the clinical performance of the M40FD-RR preliminary SIRS monitor in 29 patients (12 female, 17 male, aged 15-90 years) hospitalized at Suwa Central Hospital with COVID-19, which was confirmed by a positive polymerase chain reaction test. SIRS was evaluated by logistic regression analysis using M40FD-RR, heart rate, age, and sex as explanatory variables. We compared the results of 109 examinations of 29 COVID-19 inpatients with SIRS against those determined by the proposed monitor. The proposed monitor achieved 75% sensitivity and 83% negative predictive value, making it a promising candidate for future 24 h noninvasive preliminary SIRS tests.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11598245 | PMC |
http://dx.doi.org/10.3390/s24227100 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!