Background: The use of 3D-printing in medicine requires a context-specific quality assurance program to ensure patient safety. The process of medical 3D-printing involves several steps, each of which might be prone to its own set of errors. The segmentation error (SegE), the digital editing error (DEE) and the printing error (PrE) are the most important partial errors. Approaches to evaluate these have not yet been implemented in a joint concept. Consequently, information on the stability of the overall process is often lacking and possible process optimizations are difficult to implement. In this study, SegE, DEE, and PrE are evaluated individually, and error propagation is used to examine the cumulative effect of the partial errors.

Methods: The partial errors were analyzed employing surface deviation analyses. The effects of slice thickness, kernel, threshold, software and printers were investigated. The total error was calculated as the sum of SegE, DEE and PrE.

Results: The higher the threshold value was chosen, the smaller were the segmentation results. The deviation values varied more when the CT slices were thicker and when the threshold was more distant from a value of around -400 HU. Bone kernel-based segmentations were prone to artifact formation. The relative reduction in STL file size [as a proy for model complexity] was greater for higher levels of smoothing and thinner slice thickness of the DICOM datasets. The slice thickness had a minor effect on the surface deviation caused by smoothing, but it was affected by the level of smoothing. The PrE was mainly influenced by the adhesion of the printed part to the build plate. Based on the experiments, the total error was calculated for an optimal and a worst-case parameter configuration. Deviations of 0.0093 mm ± 0.2265 mm and 0.3494 mm ± 0.8001 mm were calculated for the total error.

Conclusions: Various parameters affecting geometric deviations in medical 3D-printing were analyzed. Especially, soft reconstruction kernels seem to be advantageous for segmentation. The concept of error propagation can contribute to a better understanding of the process specific errors and enable future analytical approaches to calculate the total error based on process parameters.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11583775PMC
http://dx.doi.org/10.1186/s41205-024-00242-xDOI Listing

Publication Analysis

Top Keywords

medical 3d-printing
12
error propagation
12
slice thickness
12
total error
12
error
9
geometric deviations
8
deviations medical
8
partial errors
8
sege dee
8
surface deviation
8

Similar Publications

Mid-infrared photoacoustic microscopy can capture biochemical information without staining. However, the long mid-infrared optical wavelengths make the spatial resolution of photoacoustic microscopy significantly poorer than that of conventional confocal fluorescence microscopy. Here, we demonstrate an explainable deep learning-based unsupervised inter-domain transformation of low-resolution unlabeled mid-infrared photoacoustic microscopy images into confocal-like virtually fluorescence-stained high-resolution images.

View Article and Find Full Text PDF

Template-Assisted Electrospinning and 3D Printing of Multilayered Hierarchical Vascular Grafts.

J Biomed Mater Res B Appl Biomater

January 2025

Department of Polymer and Biomaterials Science, Faculty of Chemical Technology and Engineering, West Pomeranian University of Technology in Szczecin, Szczecin, Poland.

Fabricating complex hierarchical structures mimicking natural vessels and arteries is pivotal for addressing problems of cardiovascular diseases. Various fabrication strategies have been explored to achieve this goal, each contributing unique advantages and challenges to the development of functional vascular grafts. In this study, a three-layered tubular structure resembling vascular grafts was fabricated using biocompatible and biodegradable copolymers of poly(butylene succinate) (PBS) using advanced manufacturing techniques.

View Article and Find Full Text PDF

Background: Treating infectious bone defects combined with large soft-tissue lesions poses significant clinical challenges. Herein, we introduced a modified two-stage treatment approach involving the implantation of 3D-printed prostheses and flap repair to treat large segmental infectious tibial bone defects.

Method: We conducted a retrospective study of 13 patients treated at our center between April 2018 and March 2022 for tibial infections owing to posttraumatic infection and chronic osteomyelitis combined with soft tissue defects.

View Article and Find Full Text PDF

Higher interfragmentary compression force improves lateral tibial plateau fracture stability using locking plate fixation: experimental and simulation verification.

BMC Musculoskelet Disord

December 2024

Department of Anatomy, Guangdong Provincial Key Laboratory of Digital Medicine and Biomechanics, Guangdong Engineering Research Center for Translation of Medical 3D Printing Application, National Virtual & Reality Experimental Education Center for Medical Morphology, School of Basic Medical Sciences, Southern Medical University, No.1023, South Shatai Road, Baiyun District, Guangzhou, Guangdong, 510515, China.

Background: This study investigated the impact of higher interfragmentary compression force (IFCF) on the stability of locking plate fixation in lateral tibial plateau fractures.

Methods: Biomechanical experiments and finite element analysis (FEA) were employed to compare the performance of the AO cancellous lag screw (AOCLS) and a newly developed combined cancellous lag screw (CCLS).

Results: The results demonstrated that the CCLS provided a higher IFCF without the risk of over-screwing, significantly improving fixation stability.

View Article and Find Full Text PDF

Background: To evaluate the feasibility, effectiveness and assistant effect of 3D printed aortic model in the treatment on congenital coarctation of the aorta (CoA) in adolescents and adults.

Methods: From December 2018 to December 2023, a total of 10 patients with congenital coarctation of aorta underwent percutaneous balloon dilatation covered stent implantation in the department of cardiovascular surgery, Xijing Hospital. There were 6 males and 4 females whose average age was (27.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!