AI Article Synopsis

  • Instability after total hip arthroplasty (THA) is a key reason for needing revision surgeries, highlighting the importance of achieving stability through different femoral systems.
  • The study used Autodesk Fusion 360 software to simulate and analyze how various head and neck sizes affect the arc of motion (AOM) and jump distances (JDs) in THA and dual mobility systems.
  • Results showed that while larger head sizes generally increase AOM, benefits plateau after 36 mm, and using modern polyethylene liners can enhance stability; thus, choosing between THA and dual mobility involves weighing minimal gains in head size against increased stability with larger heads.

Article Abstract

Background: Instability following total hip arthroplasty (THA) is a leading cause of revisions. Our objective was to evaluate the options that the surgeon has, to gain inherent stability with the use of conventional large femoral and dual mobility systems, and how the arc of motion (AOM) and jump distances (JDs) vary between them.

Methods: The head sizes examined spanned from 22 mm-54 mm, and neck sizes spanned from 10 mm-14 mm. Autodesk Fusion 360 is full-scale computer-aided designsoftware that can run simulations to validate a design. It was employed to calculate the AOM for each model. The JD was calculated with varying head sizes and 10- and 20-degree highwall liners.

Results: Increasing head sizes from 22, 28, 32-36 mm showed a considerable increase in the AOM for every neck size; however, there was substantially less of an increase with head sizes larger than 36 mm. As neck sizes increased from 10-14 mm, the AOM decreased. The JD increased substantially with the addition of 10- and 20-degree highwall liners.

Conclusions: Both classical THA and the dual mobility systems achieve greater AOM than an anatomical normal hip. As the head diameter increases, the AOM and the JD increases. Increases in head size improve range of motion; however, head sizes over 36-40 mm gain little regarding component stability. The more durable and thinner modern polyethylene liners allow for larger conventional femoral heads. When considering between a classical THA or a dual mobility system, the minimal gains with increasing the head size in using a dual mobility system may be far outweighed by the increase in directional stability offered by the large head THA with highwall liners.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11570759PMC
http://dx.doi.org/10.1016/j.artd.2024.101556DOI Listing

Publication Analysis

Top Keywords

head sizes
20
dual mobility
16
head
10
arc motion
8
total hip
8
hip arthroplasty
8
mobility systems
8
neck sizes
8
10- 20-degree
8
20-degree highwall
8

Similar Publications

Background: Understanding resilience factors in children is essential for developing early mental health interventions. Middle childhood is an understudied developmental stage, with many quantitative measures lacking validation for this age group and not capturing diverse experiences. This study aimed to use body mapping, an arts-based method, as a novel approach to understand 7-10-year-old children's concepts of resilience (including definitions and factors that contribute to resilience) in East London.

View Article and Find Full Text PDF

Subconcussive impacts are very common in the sports world and can have many negative impacts on human function, including increased risk for cognitive decline and behavioral impairments such as chronic traumatic encephalopathy (CTE). The purpose of this article is to analyze the available literature on the effects of jugular vein compression applied by a cervical collar on cerebral structure and function in the setting of chronic impact exposure. This narrative review analyzed 17 articles on brain structure and function, published between 1992 and 2022.

View Article and Find Full Text PDF

Introduction: Virtual reality (VR) holds significant promise for psychiatric research, treatment, and assessment. Its unique ability to elicit immersion and presence is important for effective interventions. Immersion and presence are influenced by matching-the alignment between provided sensory information and user feedback, and self-presentation-the depiction of a user's virtual body or limbs.

View Article and Find Full Text PDF

The Association of Jumbo Femoral Heads and Acetabular Cup Position on Patient Outcomes after Modern Posterior Approach Total Hip Arthroplasty.

J Arthroplasty

January 2025

Department of Orthopaedic Surgery, Indiana University School of Medicine, Indianapolis, Indiana; Indiana Joint Replacement Institute, Indianapolis, Indiana. Electronic address:

Background: Use of jumbo femoral heads (≥ 40 mm) in total hip arthroplasty (THA) decreases postoperative dislocation, however, may leave patients more susceptible to groin pain. Limited data exist for the effect of jumbo femoral heads and acetabular cup position on modern patient-reported outcome measures (PROMs). This study evaluated the effect of jumbo femoral heads and acetabular cup position on PROMs after primary THA.

View Article and Find Full Text PDF

Infrahyoid myocutaneous flap for reconstruction in head and neck cancer patients.

Natl J Maxillofac Surg

November 2024

Department of Radiation Oncology, Dr. RMLIMS, Lucknow, Uttar Pradesh, India.

Background: Reconstruction of head and neck defects following tumor excision is one of the most challenging surgeries due to multiple reasons, such as associated cosmetic and functional impairments. The three-dimensional aspect of the defect makes it more difficult. Although in modern settings, microvascular surgery is preferred in many conditions, it requires the presence of resources and expertise.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!