Evaluation of C/C ratio measurements using accelerator mass spectrometry with standard materials under different graphitization conditions.

Appl Radiat Isot

Department of Energy and Electrical Engineering, Dongguk University, 123 Dongdae-ro, Gyeongju-si, 38066, South Korea. Electronic address:

Published: January 2025

In this study, an evaluation of carbon-beam tuning and C/C ratio measurements is performed to validate the graphitization process in the pre-processing procedure employed at the accelerator mass spectrometry (AMS) facility of Dongguk University in Korea. The AMS isotopic ratio data are analyzed for samples subjected to three different graphitization conditions: (1) the entire process including Fe catalyst pre-heating with a high vacuum, (2) a high vacuum without pre-heating, and (3) a low vacuum without pre-heating. High-quality isotopic ratio measurements are achieved under these conditions. The measured mean isotopic ratio values for the background samples are almost equivalent to the different graphitization condition samples, with a difference in the order of 10. Therefore, conducting measurements with or without Fe catalyst pre-heating or high vacuum conditions is equally viable. Furthermore, the simplification of the graphitization process steps reduces the processing time required to produce five samples from 8 to 4 h, thereby enhancing the daily sample throughput by a factor of two, without compromising data quality.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2024.111589DOI Listing

Publication Analysis

Top Keywords

ratio measurements
12
isotopic ratio
12
high vacuum
12
c/c ratio
8
accelerator mass
8
mass spectrometry
8
graphitization conditions
8
graphitization process
8
catalyst pre-heating
8
pre-heating high
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!