Feasibility and safety of endomyocardial biopsy by lumenless pacing lead sheath-directed method during cardiac implantable electronic device implantation.

Heart Rhythm

Department of Cardiology, Sir Run Run Shaw Hospital, College of Medicine, Zhejiang University, Hangzhou, Zhejiang, People's Republic of China; Key Laboratory of Cardiovascular Intervention and Regenerative Medicine of Zhejiang Province, People's Republic of China. Electronic address:

Published: November 2024

AI Article Synopsis

Article Abstract

Background: Patients with an indication for a cardiovascular implantable electronic device (CIED) are complicated with special cardiomyopathy or other unspecified cardiac abnormalities and may need endomyocardial biopsy (EMB). However, EMB by a bioptome is usually avoided to reduce the risk of lead displacement in the CIED periprocedural period.

Objective: We aimed to assess the safety and feasibility of a novel approach for transvenous right ventricular (RV) EMB using the lead sheath method (L-S-M) during CIED implantation and compared it with the traditional bioptome method (T-B-M).

Methods: A consecutive 80 patients referred for EMB were enrolled. In the L-S-M group, a sheath with a lumenless pacing lead was positioned toward the middle to apical interventricular septum under fluoroscopy (n = 60). The CIED implantation was performed through the same venous access site. In the T-B-M group, a bioptome was used (n = 20). The clinical characteristics, procedural details, success rate, and complications were evaluated.

Results: There were 380 RV EMB procedures performed with comparable numbers of samples per patient, 4.1 ± 0.8 in the L-S-M group and 3.8 ± 0.8 in the T-B-M group. In the L-S-M group, 7 (11.7%) patients experienced minor complications with 3 transient right bundle branch blocks, 2 transient atrioventricular blocks, and 2 regional minor device pocket hematomas. In the T-B-M group, 3 (15%) experienced 1 temporary pacing for transient atrioventricular block, 1 chest discomfort, and 1 regional hematoma. No cardiac tamponade was detected.

Conclusion: RV EMB by the innovative L-S-M is technically feasible and safe. It can yield valuable and early diagnostic insights for patients who are candidates for CIEDs.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2024.11.018DOI Listing

Publication Analysis

Top Keywords

l-s-m group
12
t-b-m group
12
endomyocardial biopsy
8
lumenless pacing
8
pacing lead
8
implantable electronic
8
electronic device
8
cied implantation
8
transient atrioventricular
8
emb
6

Similar Publications

Feasibility and safety of endomyocardial biopsy by lumenless pacing lead sheath-directed method during cardiac implantable electronic device implantation.

Heart Rhythm

November 2024

Department of Cardiology, Sir Run Run Shaw Hospital, College of Medicine, Zhejiang University, Hangzhou, Zhejiang, People's Republic of China; Key Laboratory of Cardiovascular Intervention and Regenerative Medicine of Zhejiang Province, People's Republic of China. Electronic address:

Article Synopsis
View Article and Find Full Text PDF
Article Synopsis
  • * In a Phase II clinical trial, 27 patients received entinostat followed by nivolumab, resulting in an objective response rate of 11% and a median response duration of over 10 months, although the primary endpoint for overall effectiveness was not reached.
  • * The combination treatment led to significant immune profile changes, including increased dendritic cell activity and enhanced inflammatory response, suggesting potential for improving treatment strategies in PDA despite
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!