Purpose: Growing interest in microperimetry (MP) or fundus-controlled perimetry (FCP) as targeted psychometric testing method in geographic atrophy (GA) is warranted due to the disease subclinical/extra-foveal appearance or preexisting foveal loss with visual acuity becoming unreliable. We provide comprehensive pointwise test-retest repeatability reference values on the most widely used MP devices and combine them with targeted testing in areas of retinal pigment epithelium (RPE) as well as photoreceptor (PR) integrity loss, guiding the interpretation of sensitivity loss during the long-term follow-up of GA patients.

Design: Prospective reliability study METHODS: Patients with GA underwent consecutive testing on CenterVue (iCare) MAIA and NIDEK MP3 devices. Obtained PWS measurements were spatially co-registered to an optical coherence tomography (OCT) volume scan acquired during the same visit. Areas with RPE and PR integrity loss, drusen and PR thickness as well as the volume of hyperreflective foci (HRF) where identified and quantified using a set of validated deep learning-based algorithms. Test-retest repeatability was assessed according to areas defined by biomarker-specific morphologic changes using Bland-Altmann coefficients of repeatability (CoR). Furthermore, the inter-device correlation, the repeatability of scotoma point detection as well as any potential effects on fixation stability were assessed.

Results: 900 stimuli per device from twenty subjects were included. Identical overall PWS test-retest variance could be detected for MAIA (±6.57) and MP3 (±6.59). PR integrity loss was associated with a higher test-retest variance on both devices (MAIA: p=0.002; MP3: p<0.001). Higher CoR for stimuli in areas presenting RPE loss (±10.99 vs ±5.34) or HRF (±9.21 vs ±6.25) could only be detected on MP3 examinations (p<0.001 and p=0.01, respectively). An excellent intra-device correlation (MAIA: 0.94[0.93-0.95] MP3: 0.94[0.94-0.95]) and a good mean inter-device correlation (0.84[0.53-0.92]) could be demonstrated. The chosen device, run order or absence of foveal sparing had no significant effect on fixation stability.

Conclusion: Areas presenting automatically quantified PR integrity loss with and without underlying RPE loss are associated with higher test-retest variance for both MAIA and MP3. These findings are crucial for an accurate interpretation of GA progression during long-term follow-up and the planning of future trials with microperimetry testing as functional study endpoint.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2024.11.005DOI Listing

Publication Analysis

Top Keywords

integrity loss
12
areas rpe
8
geographic atrophy
8
test-retest repeatability
8
test-retest variance
8
loss
6
repeatability
5
repeatability microperimetry
4
areas
4
microperimetry areas
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!