The National Metrology Institute of Japan plan to use an inner-through-type ionization chamber (IC) as a working standard to calibrate radon monitors. Novel methods to calibrate the IC were described, and the calibration factor, CF [Bq m -3 A-1], of the IC with its expanded uncertainty (k = 2) was evaluated to be (2.0 ± 0.2) × 1016 Bq m-3 A-1 in the order of 10-13 A in the IC output current. In this study, the CFs, evaluated under sampling conditions between continuous gas flowing and stationary gas filled inside the IC, were in agreement within the uncertainty.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/rpd/ncae028 | DOI Listing |
Phys Med
January 2025
Department of Medical Physics, Apollo Proton Cancer Centre, 100 Feet Road Taramani, Chennai, Tamil Nadu, India. Electronic address:
Objectives: The purpose of this study was to investigate the fundamental properties of spot-scanning proton beams and compare them to Monte Carlo (MC) simulations, both with and without CT calibration, using spatially diverse combinations of materials.
Methods: A heterogeneous phantom was created by spatially distributing titanium, wax, and thermocol to generate six scenarios of heterogeneous combinations. Proton pencil beams ranging in energy from 100 to 226.
Phys Med Biol
December 2024
Division of Radiation Oncology, National Cancer Centre Singapore, 11 Hospital Crescent, Singapore, Singapore, 169610, SINGAPORE.
Reference dosimetry measurement in a pencil beam scanning system can exhibit dose fluctuation due to intra-spill spot positional drift. This results in a noisy reference dosimetry measurement against energy which could introduce errors in monitor unit calibration. The aim of this study is to investigate the impact of smoothing the reference dosimetry measurements on the type A uncertainty.
View Article and Find Full Text PDFMed Phys
December 2024
Division of Ionizing Radiation Metrology, National Institute of Metrology (NIM), Beijing, China.
Background: The clinical use of flattening filter free (FFF) radiotherapy has significantly increased in recent years due to its effective enhancement of dose rates and reduction of scatter dose. A proposal has been made to adjust the incident electron angle of the accelerator to expand the application of FFF beams in areas such as large planning target volumes (PTVs). However, the inherent softening characteristics and non-uniformity of lateral dose distribution in FFF beams inevitably lead to increased dosimetry errors, especially for ionization chambers widely used in clinical practice, which may result in serious accidents during FFF radiotherapy.
View Article and Find Full Text PDFAppl Radiat Isot
December 2024
Radiation Physics Division, National Institute of Standards and Technology, 100 Bureau Dr., Gaithersburg, MD, 20899-8462, USA.
The massic activity of Ac in 0.1 mol/L HCl was measured by multiple primary methods over four consistent measurement campaigns. Results from the triple-to-double coincidence ratio (TDCR) method of liquid scintillation (LS) counting were in accord with other LS-based primary methods.
View Article and Find Full Text PDFBiomed Phys Eng Express
December 2024
Medical Physics Consultant, INTECNUS Foundation, RP82 8400, San Carlos de Bariloche, Río Negro, Argentina.
. To investigate the effect of the position and orientation of the detector and its influence on the determination of output factors (OF) for small fields for a linear accelerator (MR-linac) integrated with 1.5 T magnetic resonance following the TRS-483 formalism.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!