Novel analytical approach for baclofen quantification in rodent plasma.

Biomed Chromatogr

School of Pharmaceutical Sciences, Lovely Professional University, Phagwara, Punjab, India.

Published: January 2025

A simple, accurate, and robust reverse-phase high-performance liquid chromatography (RP-HPLC) method was developed and validated for the quantification of Baclofen in rat plasma. The method demonstrated high degree of linearity (r = 0.9936) across a concentration range of 10-50 μg/mL. Precision, accuracy, limit of detection (LOD), limit of quantification (LOQ), and robustness were evaluated according to ICH guidelines. The LOD and LOQ were found to be 0.076197 and 0.23090 μg/mL, respectively. This method provides an efficient approach for Baclofen quantification in plasma, making it suitable for pharmacokinetic and bioavailability studies. The novelty of this study lies in its optimization for routine use in laboratories, ensuring reproducibility with minimal variations across different conditions and analysts.