Hybrid Reconstruction Approach for Polychromatic Computed Tomography in Highly Limited-Data Scenarios.

Sensors (Basel)

Bioengineering Department, Universidad Carlos III de Madrid, 28911 Leganes, Spain.

Published: October 2024

Conventional strategies aimed at mitigating beam-hardening artifacts in computed tomography (CT) can be categorized into two main approaches: (1) postprocessing following conventional reconstruction and (2) iterative reconstruction incorporating a beam-hardening model. While the former fails in low-dose and/or limited-data cases, the latter substantially increases computational cost. Although deep learning-based methods have been proposed for several cases of limited-data CT, few works in the literature have dealt with beam-hardening artifacts, and none have addressed the problems caused by randomly selected projections and a highly limited span. We propose the deep learning-based prior image constrained (PICDL) framework, a hybrid method used to yield CT images free from beam-hardening artifacts in different limited-data scenarios based on the combination of a modified version of the Prior Image Constrained Compressed Sensing (PICCS) algorithm that incorporates the L2 norm (L2-PICCS) with a prior image generated from a preliminary FDK reconstruction with a deep learning (DL) algorithm. The model is based on a modification of the U-Net architecture, incorporating ResNet-34 as a replacement of the original encoder. Evaluation with rodent head studies in a small-animal CT scanner showed that the proposed method was able to correct beam-hardening artifacts, recover patient contours, and compensate streak and deformation artifacts in scenarios with a limited span and a limited number of projections randomly selected. Hallucinations present in the prior image caused by the deep learning model were eliminated, while the target information was effectively recovered by the L2-PICCS algorithm.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11548251PMC
http://dx.doi.org/10.3390/s24216782DOI Listing

Publication Analysis

Top Keywords

beam-hardening artifacts
16
prior image
16
computed tomography
8
limited-data scenarios
8
deep learning-based
8
randomly selected
8
limited span
8
image constrained
8
deep learning
8
beam-hardening
5

Similar Publications

A comprehensive dual energy method for CBCT metal artifact reduction.

Phys Med Biol

December 2024

Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui 230026, People's Republic of China.

A major limitation in cone beam CT (CBCT) application is the presence of metal artifacts when scanning metal-embedded objects or high attenuation materials. This study aims to develop a dual-energy based method for effective metal artifact reduction.The proposed method comprised three steps.

View Article and Find Full Text PDF
Article Synopsis
  • * Recent literature has focused on new technologies to improve the diagnosis and prognosis of head and neck squamous cell carcinoma (HNSCC), with particular attention on dual-energy computed tomography (DECT).
  • * DECT uses two X-ray energy levels to create detailed images that help differentiate materials, improve contrast, and assist in tumor assessment, potentially offering significant advantages in managing HNSCC.
View Article and Find Full Text PDF

Conventional strategies aimed at mitigating beam-hardening artifacts in computed tomography (CT) can be categorized into two main approaches: (1) postprocessing following conventional reconstruction and (2) iterative reconstruction incorporating a beam-hardening model. While the former fails in low-dose and/or limited-data cases, the latter substantially increases computational cost. Although deep learning-based methods have been proposed for several cases of limited-data CT, few works in the literature have dealt with beam-hardening artifacts, and none have addressed the problems caused by randomly selected projections and a highly limited span.

View Article and Find Full Text PDF

Study on the impact of bowtie filter on photon-counting CT imaging.

Phys Med Biol

October 2024

Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, Shenzhen, Guangdong 518055, People's Republic of China.

The aim of this study was to investigate the impact of the bowtie filter on the image quality of the photon-counting detector (PCD) based CT imaging.Numerical simulations were conducted to investigate the impact of bowtie filters on image uniformity using two water phantoms, with tube potentials ranging from 60 to 140 kVp with a step of 5 kVp. Subsequently, benchtop PCD-CT imaging experiments were performed to verify the observations from the numerical simulations.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to determine if the improved urate analysis (IUA) algorithm can lower false positives in CT scans of gout compared to the current algorithm (CUA).
  • Conducted over a year, it involved comparing spectral CT images of suspected gout cases using both algorithms against a reference standard (ultrasound + conventional CT).
  • Results showed that IUA significantly improved specificity and accuracy while also reducing image artifacts compared to CUA, although the maximum area of tophi was measured to be greater in CUA images.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!