Background: The clinical potential of Raman spectroscopy is well established but has yet to become established in routine oncology workflows. One barrier slowing clinical adoption is a lack of evidence demonstrating that data taken on one spectrometer transfers across to data taken on another spectrometer to provide consistent diagnoses.
Methods: We investigated multi-centre transferability using human oesophageal tissue. Raman spectra were taken across three different centres with different spectrometers of the same make and model. By using a common protocol, we aimed to minimise the difference in machine learning performance between centres.
Results: 61 oesophageal samples from 51 patients were interrogated by Raman spectroscopy at each centre and classified into one of five pathologies. The overall accuracy and log-loss did not significantly vary when a model trained upon data from any one centre was applied to data taken at the other centres. Computational methods to correct for the data during pre-processing were not needed.
Conclusion: We have found that when using the same make and model of spectrometer, together with a common protocol, across different centres it is possible to achieve system transferability without the need for additional computational instrument correction.