The study aims to evaluate the energy response of the thermoluminescent dosemeter (TLD) CaSO4:Dy from high energy photon beams produced from medical linear accelerator. The test was performed on the polymethyl methacrylate phantom surface and at the depth of dose maximum with a source-to-surface distance of 100 cm and the radiation field size of 12 × 12 cm. The results were compared with the TLD response exposed to 60Co standard source. The results show that the TLD response to the linac's photon differs >30% compared to the TLD response to 60Co exposure when measured at the phantom surface, while at the depth of maximum dose the response is similar (< ±10%), indicating that the TLD response is no longer dependent on the energy. This implies that the current dose reading evaluation based on 60Co calibration curve should be corrected to obtain more accurate dose report of the radiation workers.