AI Article Synopsis

  • The peQasus study evaluates the safety and effectiveness of a new temperature-controlled high-power short-duration (HPSD) radiofrequency catheter (QDOT Micro) for pulmonary vein isolation (PVI) in patients with atrial fibrillation (AF).
  • Conducted across 15 centers in Europe, the study involved 1,023 patients and compared outcomes between a very HPSD technique and a hybrid approach.
  • Results indicated complete PVI success, a mean procedure time of about 98 minutes, and a similar safety profile for both techniques, with 12-month arrhythmia-free survival rates around 77%.

Article Abstract

Background: Temperature-controlled high-power short-duration (HPSD) radiofrequency catheter ablation for pulmonary vein isolation (PVI) utilizing a novel ablation catheter (QDOT Micro) with real-time assessment of catheter tip temperature aims for safer, more effective and faster procedures.

Methods: The peQasus study is a large European multicenter study set up to assess safety, acute efficacy and outcomes of temperature-controlled HPSD based PVI. The primary endpoints were safety, efficacy and 12-months freedom from atrial tachyarrhythmias. Additionally, two strategies namely very HPSD (90W for 4 seconds) only and a hybrid approach (HPSD with maximum of 50W and vHPSD) were compared.

Results: A total of 1,023 AF patients in 15 centers from 9 European countries received PVI with the QDOT. Complete PVI was successfully achieved in all patients. In 699/1023 (68.3%) the vHPSD only approach (vHPSD group) and in 324/(31.7%) patients the hybrid approach (hybrid group) was utilized. The mean procedure duration was 98.4±37.4 min (vHPSD: 88.2±34.9min, hybrid: 117.4±32.7min, p<0.001). The first pass isolation rate of all PVs was 64% (vHPSD: 62.6%, hybrid: 67.1%, p=0.187). Severe adverse events were observed in 1.7% (vHPSD: 1.6%, hybrid: 1.9%, p=0.746). 12-month arrhythmia-recurrence free survival was 77.1% (vHPSD: 76.8%, hybrid: 77.8%, p=0.241).

Conclusions: In this large multicentre study temperature-controlled HPSD and vHPSD ablation via a novel ablation catheter provides safe and effective PVI with a relatively short procedure duration. Despite a shorter procedure time no differences in terms of safety and freedom from arrhythmia-recurrence were found irrespective of utilizing vHPSD or the hybrid approach.

Download full-text PDF

Source
http://dx.doi.org/10.1093/europace/euae284DOI Listing

Publication Analysis

Top Keywords

hybrid approach
8
temperature guided
4
guided high
4
high high-power
4
high-power short
4
short duration
4
duration ablation
4
ablation atrial
4
atrial fibrillation
4
fibrillation treatment
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!