Background: Continuous positive airway pressure (CPAP) is a recommended first-line therapy for infants with respiratory distress at birth. Resuscitation devices incorporating CPAP delivery can have significantly different imposed resistances affecting airway pressure stability and work of breathing.
Aim: To compare CPAP performance of two resuscitation devices (Neopuff T-piece resuscitator and rPAP) in a neonatal lung model simulating spontaneous breathing effort at birth.
Methods: The parameters assessed were variation in delivered pressures (∆P), tidal volume (VT), inspiratory effort (model pressure respiratory muscle (PRM)) and work of breathing (WOB). Two data sequences were required with Neopuff and one with rPAP: (1) set PRM with changes in VT and (2) constant VT (preterm 6 mL, term 22 mL) with increased effort. Data were collected at CPAP settings of 5, 7 and 9 cmHO using a 1 kg preterm (Compliance: 0.5 mL/cmHO) and 3.5 kg term (1.0 mL/cmHO) model.
Results: 2298 breaths were analysed (760 rPAP, 795 Neopuff constant VT, 743 Neopuff constant PRM). With CPAP at 9 cmHO and set VT the mean ∆P (cmHO) rPAP vs Neopuff 1.1 vs 5.6 (preterm) and 1.9 vs 13.4 (term), WOB (mJ) 4.6 vs 6.1 (preterm) and 35.3 vs 44.5 (term), and with set PRM mean VT (ml) decreased to 6.2 vs 5.2 (preterm) and 22.3 vs 17.5 (term) p<0.001. Similar results were found at pressures of 5 and 7 cmHO.
Conclusion: rPAP had smaller pressure swings than Neopuff at all CPAP levels and was thus more pressure stable. WOB was higher with Neopuff when VT was held constant. VT reduced with Neopuff when respiratory effort was constant.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11552556 | PMC |
http://dx.doi.org/10.1136/bmjpo-2024-002948 | DOI Listing |
Early Hum Dev
December 2024
Department of Neonatology, Wilhelmina Children's Hospital, University Medical Center Utrecht, Lundlaan 6, 3584 EA Utrecht, the Netherlands.
Introduction: Non-invasive respiratory support strategies have evolved to avoid bronchopulmonary dysplasia (BPD) in preterm infants. However, consensus on the best treatment strategy remains lacking. This study aims to investigate current practices and variations in primary respiratory support for extremely preterm neonates across neonatal intensive care units (NICUs) in the Netherlands.
View Article and Find Full Text PDFResuscitation
December 2024
Neonatal Unit, National Maternity Hospital, Dublin, Ireland; School of Medicine, University College Dublin, Ireland. Electronic address:
Background: Most preterm infants breathe spontaneously at birth. Despite this, the majority have a face mask immediately applied for breathing support. Face mask application may inhibit spontaneous breathing in newborn infants.
View Article and Find Full Text PDFPediatrics
January 2025
Virginia Mason Franciscan Health, Tacoma, Washington.
Objective: Our institutional data revealed high pneumothorax rates in term neonates resuscitated in the delivery room (DR). Other studies have reported that high rates of continuous positive airway pressure (CPAP) in the DR are associated with increased pneumothorax rates. We sought to test the hypothesis that quality improvement efforts to reduce the use of CPAP in the DR would be associated with a reduced incidence of pneumothorax.
View Article and Find Full Text PDFAntioxidants (Basel)
October 2024
Department of Pediatrics, Joan C. Edwards School of Medicine, Marshall University, Huntington, WV 25701, USA.
Apnea of prematurity (AOP) occurs in 85% of neonates ≤34 weeks of gestational age. AOP is frequently associated with intermittent hypoxia (IH). This narrative review reports on the putative relationship of AOP with IH and the resulting oxidative stress (OS).
View Article and Find Full Text PDFCrit Care
November 2024
Section of Anaesthetics, Pain Medicine and Intensive Care, Department of Surgery and Cancer, Faculty of Medicine, Imperial College London, London, UK.
Background: High flow nasal cannula therapy (HFNC) and continuous positive airway pressure (CPAP) are two widely used modes of non-invasive respiratory support in paediatric critical care units. The FIRST-ABC randomised controlled trials (RCTs) evaluated the clinical and cost-effectiveness of HFNC compared with CPAP in two distinct critical care populations: acutely ill children ('step-up' RCT) and extubated children ('step-down' RCT). Clinical effectiveness findings (time to liberation from all forms of respiratory support) showed that HFNC was non-inferior to CPAP in the step-up RCT, but failed to meet non-inferiority criteria in the step-down RCT.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!