A New Intraocular Lens Power Formula Integrating an Artificial Intelligence-Powered Estimation for Effective Lens Position Based on Chinese Eyes.

Transl Vis Sci Technol

Department of Ophthalmology, Shanghai East Hospital, Tongji University School of Medicine, Tongji University, Shanghai, China.

Published: October 2024

Purpose: To develop and evaluate a new intraocular lens (IOL) formula based on Chinese eyes.

Methods: A training dataset of 709 eyes undergoing uneventful cataract surgery was used to train the algorithm for effective lens position estimation. The algorithm was then integrated with Gaussian optics to develop the new IOL formula (Jin-AI). From the same center, 177 eyes served as an internal test dataset. An independent dataset of 557 eyes served as an external test dataset. The standard deviation (SD) of prediction errors was compared among the Jin-AI formula, traditional third-generation formulas (SRK/T, Holladay 1, Hoffer Q), and newer generation formulas (Kane, Barrett Universal II [BUII], Hill-radial basis function [RBF] 3.0, and PEARL-DGS).

Results: In the internal test dataset, the Jin-AI formula showed the lowest SD (0.25 D), followed by the BUII (0.31 D), Kane (0.33 D), and PEARL-DGS (0.33 D) formulas. In the external test dataset, the Jin-AI, Kane, and PEARL-DGS formulas had the lowest SD (0.38 D), followed by the BUII (0.39 D), Hill-RBF 3.0 (0.39 D), SRK/T (0.45 D), Holladay 1 (0.48 D), and Hoffer Q (0.48 D) formulas. The SD of the Jin-AI formula was significantly lower than the third-generation formulas and comparable to the four newer generation formulas. Predictive accuracy of the Jin-AI formula was similar to the newer generation formulas across all axial length, keratometry, and anterior chamber depth ranges.

Conclusions: The new formula has exhibited good performance in predicting postoperative refractions. Its overall predictive accuracy was better than the third-generation formulas and comparable to the newer generation ones.

Translational Relevance: The Jin-AI formula could be a reliable alternative for IOL power calculation in Chinese.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11534016PMC
http://dx.doi.org/10.1167/tvst.13.10.40DOI Listing

Publication Analysis

Top Keywords

jin-ai formula
20
test dataset
16
newer generation
16
third-generation formulas
12
generation formulas
12
formula
9
formulas
9
intraocular lens
8
effective lens
8
lens position
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!