Anti-AAV neutralizing Abs (NAbs) titer is usually measured by cell-based microneutralization (MN) assay and is crucial for patient screening in AAV-based gene therapy clinical trials. However, achieving uniform operation and comparable results among different laboratories remains challenging. Here, we established a standardized MN assay for anti-AAV9 NAbs in human sera or plasma and transferred the method to the other two research teams. Then, we validated its parameters and tested a set of eight human samples in blind across all laboratories. The end-point titer, defined by a transduction inhibition of 50% (IC), was calculated using curve-fit modelling. A mouse neutralizing monoclonal antibody in human negative serum was used for system quality control (QC), requiring inter-assay titer variation of <4-fold difference or geometric coefficient of variation (%GCV) of <50%. The assay demonstrated a sensitivity of 54 ng/mL and no cross-reactivity to 20 μg/mL anti-AAV8 MoAb. The intra-assay and inter-assay variation for the low positive QC were 7-35% and 22-41%, respectively. The titers of the blind samples showed excellent reproducibility within and among laboratories, with a %GCV of 18-59% and 23-46%, respectively. This study provides a commonly transferrable MN assay for evaluating anti-AAV9 NAbs in humans, supporting its application in clinical trials.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11512302 | PMC |
http://dx.doi.org/10.3390/v16101512 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!