Purpose: Three-dimensional imaging has become an increasingly important component of orthodontics. Associated with this, however, is a higher radiation exposure for patients. New cone-beam computed tomography (CBCT) devices have been developed that can provide low-dose CBCT (LD-CBCT). We hypothesized that LD-CBCT is as precise and reproducible as standard high-dose CBCT (HD-CBCT) in segmenting roots and crowns as well as measuring tooth length.
Methods: HD-CBCT and LD-CBCT scans were taken of four human cadaveric heads. Thirty single-rooted teeth were segmented twice by one investigator. The length of each tooth was also measured. Lin's concordance correlation coefficient (CCC) was calculated to assess the agreement of HD-CBCT and LD-CBCT measurements and the intraclass correlation coefficient (ICC) was calculated to assess intrarater reliability. Analyses were supported by Bland-Altman plots.
Results: Volume measurements obtained using HD-CBCT were significantly higher than those obtained using LD-CBCT (p < 0.001). CCC was 0.975 (95% confidence interval [CI] = 0.956-0.986) indicating excellent agreement between the two modalities. Intrarater reliability between the two sets of LD-CBCT and HD-CBCT volume measurements was excellent (ICC = 0.998, 95%CI = 0.995-0.999 [HD-CBCT], ICC = 0.997, 95%CI = 0.992-0.998 [LD-CBCT]). CCC for tooth length measurements was 0.991 (95% CI = 0.983-0.995), indicating excellent agreement between HD-CBCT and LD-CBCT. Intrarater reliabilities between the two sets of tooth length measurements were also excellent for both methods (ICC = 0.998, 95%CI = 0.995-0.999 [HD-CBCT], ICC = 0.997, 95%CI = 0.992-0.998 [LD-CBCT]).
Conclusions: Within the limitations of this experimental setting, LD-CBCT is as valid as HD-CBCT for measuring tooth length. Regarding the volume differences, in vivo studies are required to determine their clinical relevance.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00056-024-00558-7 | DOI Listing |
J Orofac Orthop
October 2024
Department of Orthodontics, University Hospital Schleswig Holstein, Arnold-Heller-Straße 3, 24105, Kiel, Germany.
Purpose: Three-dimensional imaging has become an increasingly important component of orthodontics. Associated with this, however, is a higher radiation exposure for patients. New cone-beam computed tomography (CBCT) devices have been developed that can provide low-dose CBCT (LD-CBCT).
View Article and Find Full Text PDFJ Imaging
July 2024
Department of Orthodontics, University Hospital Schleswig Holstein, Arnold-Heller-Straße 3, 24105 Kiel, Germany.
Acta Odontol Scand
April 2024
Department of Orthodontics, University Hospital Schleswig Holstein, Kiel, Germany.
Objective: The current study explores whether there is a clinically relevant distinction in the measurement of marginal bone loss when comparing high-dose (HD) versus low-dose (LD) cone beam computed tomography (CBCT) protocols in small and large acquisition volumes. Material and Methods: CBCTs of four human cadaveric preparates were taken in HD and LD mode in two different fields of view 8 × 8 cm2 (LV) and 5 × 5 cm2 (SV). In total, 43 sites of 15 teeth were randomly chosen, and marginal bone loss was measured twice in all protocols at 43 sites of 15 teeth by one calibrated investigator.
View Article and Find Full Text PDFBMC Oral Health
December 2022
Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Ain-Shams University, Cairo, Egypt.
Background: Early accurate radiographic assessment of peri-implant bone condition is highly important to avoid excessive loss of supporting bone and implant failure. Cone beam computed tomography (CBCT) is the radiographic technique of choice if peri-implant dehiscence and fenestration defects are suspected. The higher radiation dose and the presence of beam hardening artifacts are the main drawbacks of CBCT imaging techniques.
View Article and Find Full Text PDFSci Rep
April 2022
Department of Orthodontics, University Hospital Schleswig Holstein, Arnold-Heller-Straße 3, 24105, Kiel, Germany.
Different cone beam computed tomography (CBCT) protocols have shown promising results for imaging furcation defects. This study evaluates the suitability of low-dose (LD)-CBCT for this purpose. Fifty-nine furcation defects of nine upper and 16 lower molars in six human cadavers were measured by a high-dose (HD)-CBCT protocol, a LD-CBCT protocol, and a surgical protocol.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!