  • - The Aveir DR system is a new type of dual-chamber leadless pacemaker that features separate atrial and ventricular components capable of communicating wirelessly for synchronized heart pacing.
  • - A study involving 399 patients showed that the success rates of implant-to-implant (i2i) communication between the atrial and ventricular pacemakers exceeded 90% at all tested intervals over six months.
  • - Improvements in communication success were noted, particularly for patients with initial lower success rates, suggesting benefits from reprogramming and device stabilization over time.

Article Abstract

Background: Aveir DR (Abbott, Abbott Park, IL) is a dual-chamber leadless pacemaker (LP) system with distinct atrial and ventricular LPs (ALP, VLP) that communicate wirelessly to deliver atrioventricular synchronous pacing. Success rates of these implant-to-implant (i2i) transmissions have not been systematically evaluated.

Objective: This study aims to evaluate chronic i2i success rates in a clinical setting.

Methods: Patients meeting standard dual-chamber pacing indications were enrolled and implanted with dual-chamber LP systems as part of a prospective international clinical trial (Aveir DR i2i Study). The percent of successful i2i transmissions from ALP-to-VLP (A-to-V) and VLP-to-ALP (V-to-A) were interrogated from LPs in de novo patients using the device programmer at implant, discharge, and at 1, 3, and 6 months postimplant (1M, 3M, 6M).

Results: A total of 399 patients completed device implant and i2i diagnostic interrogation (62% male; age 69 years; 65% sinus node dysfunction, 32% atrioventricular [AV] block). Median A-to-V and V-to-A i2i success rates exceeded 90% of beats at all time-points from implant to 6M. The minority of patients with A-to-V or V-to-A i2i success in <70% of beats at implant (A-to-V: 19% of patients, V-to-A: 31% of patients) showed roughly 40% improvement by 1M, with this minority dropping to roughly 5% of patients by 6M. Improvement in i2i communication success may be attributed to reprogramming of i2i setting levels, natural changes in dominant posture, and device stabilization.

Conclusion: Wireless implant-to-implant communication in the new dual-chamber leadless pacemaker system demonstrated successful transmissions in >90% of beats throughout the 6-month evaluation period. Communication success improved significantly over time postimplant for specific subgroups.

Clinical Trial Registration: Aveir DR i2i Study, ClinicalTrials.gov ID NCT05252702.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2024.10.020DOI Listing

Publication Analysis

Top Keywords

success rates
12
i2i success
12
dual-chamber leadless
8
leadless pacemaker
8
i2i
8
i2i transmissions
8
aveir i2i
8
i2i study
8
a-to-v v-to-a
8
v-to-a i2i
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!