Objective: To evaluate the influence of battery level on power (mW), emission spectrum (mW/cm2/ nm), and light distribution on the active tip (mW/ cm2) of certified (FDA/ANVISA) and low-cost uncertified light-curing units (LCUs) purchased through e-commerce.
Methods: Seven LCUs, three certified: VALO Grand (Ultradent); Radii Xpert (SDI); and LED.B (Woodpecker); and four uncertified: 1 Sec; BS 300; LED curing light; and VAFU (VRN, AZDENT), were used. The LCUs were evaluated by calculating the power (mW) after each sequential five exposure cycles of 20 seconds and the emission spectrum (mW/cm2/nm) in the initial and final cycles, using an integrating sphere during three battery charging cycles. Beam profiling was used to check the light distribution on the LCU tip after every 50 exposure cycles until the battery fully discharged. Data were analyzed by linear regression between power and the number of exposure times (R2).
Results: The certified LCUs VALO Grand (R2=0.005), LED.B (R2=0.02), and Radii Xpert (R2=0.09) and the uncertified LCU VAFU (R2=0.002) had no significant power reduction during the three battery charging cycles. The uncertified LCUs BS 300 (R2=0.87), 1 Sec (R2=0.60), and LED curing light (R2=0.83) showed significant power reduction, decreasing the emission spectrum (mW/cm2/nm) at the end of the battery charging cycle. The light distribution on the active tip across the level battery was modified significantly with successive exposure times.
Conclusions: The certified LCUs (VALO Grand, Radii Xpert, and LED.B) and uncertified LCU (VAFU), maintained power, emission spectrum, and light distributions during the tested battery life cycles. Low-cost certified LCU LED.B exhibited inhomogeneous light concentrated at the center of the tip. Low-cost uncertified LCUs-BS 300, 1 Sec, and LED curing light-had significant power reductions during the battery cycles and increased inhomogeneous light distribution along the successive exposure times.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.2341/23-177-L | DOI Listing |
Phys Chem Chem Phys
January 2025
Physics Department, Khalifa University, Abu-Dhabi, United Arab Emirates.
The spectrum of carbon monoxide is important for astrophysical media, such as planetary atmospheres, interstellar space, exoplanetary and stellar atmospheres; it also important in plasma physics, laser physics and combustion. Interpreting its spectral signature requires a deep and thorough understanding of its absorption and emission properties. A new accurate spectroscopic model for the ground and electronically-excited states of the CO molecule computed at the aug-cc-pV5Z CASSCF/MRCI+Q level is reported.
View Article and Find Full Text PDFBiomed Opt Express
January 2025
Key Laboratory of Optoelectronic Devices and Systems of Ministry of Education and Guangdong Province, College of Physics and Optoelectronic Engineering, Shenzhen University, Shenzhen, China.
Multiphoton fluorescence microscopy (MFM), renowned for its noninvasiveness and high spatiotemporal resolution, is extensively applied in brain structure imaging in vivo. Three-photon fluorescence (3PF) imaging, excited at the NIR-III window, can penetrate the deepest mouse cerebrovascular. Evans blue, a substance known for its low toxicity, high water solubility, and resistance to metabolism, is frequently employed to assess blood-brain barrier (BBB) permeability.
View Article and Find Full Text PDFTerahertz (THz) generation via photomixing on photoconductive antenna using twin delayed chirped pulses provides a long THz pulse with a narrow bandwidth. To generate a long pulse with a broad bandwidth, we propose a new, to the best of our knowledge, method that combines two long optical pulses with opposite chirps. The pulses exhibit temporal distributions of their instantaneous frequencies with opposite slopes.
View Article and Find Full Text PDFRadiat Environ Biophys
January 2025
Laboratory of Health Sciences and Technologies, Higher Institute of Health Sciences, Hassan First University, Settat, Morocco.
This study assesses radiation doses in multi-slice computed tomography (CT) using epoxy resin and PMMA phantoms, focusing on the relationship between TAR (tissue air ratio) and kilovoltage peak (kVp). The research was conducted using a Hitachi Supria 16-slice CT scanner. An epoxy resin phantom was fabricated from commercially available materials, to simulate human tissue.
View Article and Find Full Text PDFNano Lett
January 2025
State Key Laboratory of Infrared Physics, Shanghai Institute of Technical Physics, Chinese Academy of Sciences, 500 Yu Tian Road, Shanghai 200083, China.
Enhancing photoluminescence (PL) efficiency in colloidal quantum dots is pivotal for next-generation near-infrared photodetectors, imaging systems, and photonic devices. Conventional methods, especially metal-based plasmonic structures, suffer from large optical losses, which limits their practical use. Here, we introduce a quasi-bound state in the continuum (quasi-BIC) metasurface on a silicon-on-insulator platform, tailored to provide high-quality factor resonances with minimized losses.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!