Acoustic noise in a small-format 3.0-T neonatal MRI system.

Pediatr Radiol

Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Ave., Cincinnati, OH, 45229, USA.

Published: November 2024

AI Article Synopsis

  • The use of MRI for premature neonates is limited due to challenges in moving them between facilities and concerns about loud scanner noise.
  • A new small-format 3.0-T MRI scanner placed in the NICU aims to overcome these issues by reducing the acoustic noise levels during imaging.
  • Results show that the NICU scanner is significantly quieter than traditional adult-sized MRI scanners, enhancing the safety and comfort for neonatal patients and staff.

Article Abstract

Background: Today, magnetic resonance imaging (MRI) is rarely used in managing the care of premature neonates. This is in large part due to the medical and logistical challenges associated with moving neonates from the neonatal intensive care unit (NICU) to the radiology department. Furthermore, acoustic noise associated with MR scanning poses safety concerns for both practitioners and neonatal patients. A small-format 3.0-T neonatal scanner was recently developed and placed within the NICU to address these logistical and acoustic challenges.

Objective: To compare acoustic noise measurements of a small-format 3.0-T neonatal MRI scanner with conventional adult-sized 1.5-T and 3.0-T MRI scanners using identical neonatal head imaging protocols.

Materials And Methods: Sound pressure level (SPL) measurements of a standard imaging protocol were made in a small-format neonatal 3.0-T MRI scanner as well as in adult-sized 1.5-T and 3.0-T scanners. SPL measurements were made with a Brüel & Kjær sound level meter model 2250. The statistical significance of the differences in SPL between scanners was determined using one-way ANOVA.

Results: Average sound pressure level values were measured in unweighted decibels (dB) and A-weighted decibels (dBA) for all imaging sequences in the protocol. The average A-weighted SPLs for the NICU from 1.5-T and 3.0-T MRI scanners were 81.02 ± 0.28 dBA, 87.00 ± 0.85 dBA, and 94.91 ± 0.65 dBA, respectively. SPLs at the isocenter of the NICU MRI scanner were 5.98 dBA quieter than in the 1.5-T scanner (P=0.007), and 13.89 dBA quieter than in the 3.0-T scanner (P<0.001). For staff standing next to the scanner, the NICU scanner was 20.24 dBA quieter than the 1.5-T scanner (P<0.001) and 19.28 dBA quieter than the 3.0-T scanner (P<0.001).

Conclusion: The NICU 3.0-T MRI system is significantly quieter than conventional adult-sized MRI systems, improving safety for neonatal patients. Significant reductions in SPL were also noted inside the screen room where clinicians may be present during scanning.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00247-024-06070-9DOI Listing

Publication Analysis

Top Keywords

acoustic noise
12
small-format 30-t
12
30-t neonatal
12
mri scanner
12
15-t 30-t
12
30-t mri
12
30-t
8
neonatal mri
8
adult-sized 15-t
8
mri scanners
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!