Reduced auditory perception and brain response with quiet TMS coil.

Brain Stimul

Department of Psychiatry and Behavioral Sciences, Duke University School of Medicine, USA; Department of Electrical and Computer Engineering, Duke University, USA; Department of Neurosurgery, Duke University School of Medicine, USA; Department of Biomedical Engineering, Duke University, USA. Electronic address:

Published: December 2024

Background: Electromagnetic forces in transcranial magnetic stimulation (TMS) coils generate a loud clicking sound that produces confounding auditory activation and is potentially hazardous to hearing. To reduce this noise while maintaining stimulation efficiency similar to conventional TMS coils, we previously developed a quiet TMS double containment coil (qTMS-DCC).

Objective: To compare the stimulation strength, perceived loudness, and EEG response between qTMS-DCC and a commercial TMS coil.

Methods: Nine healthy volunteers participated in a within-subject study design. The resting motor thresholds (RMTs) for qTMS-DCC and MagVenture Cool-B65 were measured. Psychoacoustic titration matched the Cool-B65 loudness to qTMS-DCC pulsed at 80, 100, and 120 % RMT. Event-related potentials (ERPs) were recorded for both coils. The psychoacoustic titration and ERPs were acquired with the coils both on and 6 cm off the scalp, the latter isolating the effects of airborne auditory stimulation from body sound and electromagnetic stimulation. The ERP comparisons focused on a centro-frontal region that encompassed peak responses in the global signal while stimulating the primary motor cortex.

Results: RMT did not differ significantly between the coils, with or without the EEG cap on the head. qTMS-DCC was perceived to be substantially quieter than Cool-B65. For example, qTMS-DCC at 100 % coil-specific RMT sounded like Cool-B65 at 34 % RMT. The general ERP waveform and topography were similar between the two coils, as were early-latency components, indicating comparable electromagnetic brain stimulation in the on-scalp condition. qTMS- DCC had a significantly smaller P180 component in both on-scalp and off-scalp conditions, supporting reduced auditory activation.

Conclusions: The stimulation efficiency of qTMS-DCC matched Cool-B65 while having substantially lower perceived loudness and auditory-evoked potentials.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11670816PMC
http://dx.doi.org/10.1016/j.brs.2024.10.003DOI Listing

Publication Analysis

Top Keywords

reduced auditory
8
quiet tms
8
tms coils
8
stimulation efficiency
8
perceived loudness
8
psychoacoustic titration
8
matched cool-b65
8
stimulation
7
coils
6
qtms-dcc
6

Similar Publications

The goal of the present investigation was to perform a registered replication of Jones and Macken's (1995b) study, which showed that the segregation of a sequence of sounds to distinct locations reduced the disruptive effect on serial recall. Thereby, it postulated an intriguing connection between auditory stream segregation and the cognitive mechanisms underlying the irrelevant speech effect. Specifically, it was found that a sequence of changing utterances was less disruptive in stereophonic presentation, allowing each auditory object (letters) to be allocated to a unique location (right ear, left ear, center), compared to when the same sounds were played monophonically.

View Article and Find Full Text PDF

Can one shift attention among voices at a cocktail party during a silent pause? Researchers have required participants to attend to one of two simultaneous voices - cued by its gender or location. Switching the target gender or location has resulted in a performance 'switch cost' - which was recently shown to reduce with preparation when a gender cue was presented in advance. The current study asks if preparation for a switch is also effective when a voice is selected by location.

View Article and Find Full Text PDF

Assessment of Speech Processing and Listening Effort Associated With Speech-on-Speech Masking Using the Visual World Paradigm and Pupillometry.

Trends Hear

January 2025

Faculty of Medicine and University Hospital Cologne, Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Cologne, Germany.

Speech-on-speech masking is a common and challenging situation in everyday verbal communication. The ability to segregate competing auditory streams is a necessary requirement for focusing attention on the target speech. The Visual World Paradigm (VWP) provides insight into speech processing by capturing gaze fixations on visually presented icons that reflect the speech signal.

View Article and Find Full Text PDF

Neural processing of auditory stimuli in rats: translational aspects using auditory oddball paradigms.

Behav Brain Res

January 2025

Department of Neurosurgery, Hannover Medical School, Carl-Neuberg-Straße 1, 30625 Hannover, Germany; Cluster of Excellence Hearing4all, German Research Foundation, Hannover, Germany; Center for Systems Neuroscience (ZSN) Hannover, 30559 Hannover, Germany.

Background: The three-class oddball paradigm allows to investigate the processing of behaviorally relevant and irrelevant auditory stimuli. In humans, event-related potentials (ERPs) are used as neural correlate of behavior. We recorded local field potentials (LFPs) within the medial prefrontal cortex (mPFC) in rats during three-class and passive two-class oddball paradigms and analyzed the ERPs focusing on similarities to human recordings.

View Article and Find Full Text PDF

Exploring the Hearing Improvement and Parental Stress in Children with Hearing Loss Using Hearing Aids or Cochlear Implants.

J Clin Med

December 2024

Unit of Otorhinolaryngology, Department of Adult and Development Age Human Pathology "Gaetano Barresi", University of Messina, 98122 Messina, Italy.

This study aims to describe the stress levels experienced by parents of children with hearing loss who use conventional hearing aids or cochlear implants, and to assess the correlation between parental stress and the auditory skills acquired by the children. The study was conducted at the Policlinic "Gaetano Martino" in Messina, evaluating data from 42 pairs of parents of children using hearing aids or cochlear implants. Parents completed the LittlEARS Auditory Questionnaire (LEAQ) and the Parental Stress Scale (PSS) 18 months after the initial device (hearing aid or cochlear implant) had been activated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!