The importance of factorial design on the optimization of biosensor performance: immobilization of glucose oxidase as a case study.

Anal Bioanal Chem

División de Investigación y Posgrado, Facultad de Ingeniería, Universidad Autónoma de Querétaro, 76010, Santiago de Querétaro, Mexico.

Published: December 2024

AI Article Synopsis

  • The study explored how different concentrations of glucose oxidase (GOx), ferrocene methanol (Fc), and multi-walled carbon nanotubes (MWCNTs) interact in glucose biosensors using a factorial design approach.
  • The statistical analysis revealed that GOx, MWCNT, Fc, and their interactions significantly influence the electrochemical response of the biosensors.
  • The optimal conditions identified for improved amperometric response in glucose oxidation were 10 mM GOx, 2 mg/mL Fc, and 15 mg/mL MWCNT, enhancing enzyme immobilization strategies for glucose biosensors.

Article Abstract

Conventionally, the optimization of glucose biosensors is achieved by varying the concentrations of the individual reagents used to immobilize the enzyme. In this work, the effect and interaction between glucose oxidase enzyme (GOx), ferrocene methanol (Fc), and multi-walled carbon nanotubes (MWCNTs) at different concentrations were investigated by a design of experiments (DoE). For this analysis, a factorial design with three factors and two levels each was used with the software RStudio for statistical analysis. The data were obtained by electrochemical experiments on the immobilization of GOx-Fc/MWCNT at different concentrations. The results showed that the factorial DoE method was confirmed by the non-normality of the residuals and the outliers of the experiment. When examining the effects of the variables, analyzing the half-normal distribution and the effects and contrasts for GOx-Fc/MWCNT, the factors that showed the greatest influence on the electrochemical response were GOx, MWCNT, Fc, and MWCNT:Fc, and there is a high correlation between the factors GOx, MWCNT, Fc, and MWCNT:Fc, as shown by the analysis of homoscedasticity and multicollinearity. With these statistical analyses and experimental designs, it was possible to find the optimal conditions for different factors: 10 mM mL GOx, 2 mg mL Fc, and 15 mg mL MWCNT show a greater amperometric response in the glucose oxidation. This work contributes to advancing enzyme immobilization strategies for glucose biosensor applications. Systematic investigation of DoE leads to optimized immobilization for GOx, enables better performance as a glucose biosensor, and allows the prediction of some outcomes.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00216-024-05582-5DOI Listing

Publication Analysis

Top Keywords

factorial design
8
glucose oxidase
8
gox mwcnt
8
mwcnt mwcntfc
8
glucose biosensor
8
glucose
6
gox
5
design optimization
4
optimization biosensor
4
biosensor performance
4