A fluid temperature interpolation method for wide temperature range distributed flow calorimeter.

Rev Sci Instrum

College of Metrology Measurement and Instrument, China Jiliang University, Xueyuan Street, No. 258, 310018 Hangzhou, China.

Published: October 2024

AI Article Synopsis

  • - The distributed flow calorimeter measures key thermodynamic parameters like reaction and mixing enthalpy for hydrocarbon fuels, which are essential for hypersonic flight development.
  • - A new method is proposed to calculate fluid temperature in tubes, addressing difficulties with temperature monitoring at high temperatures by interpolating the enthalpy function through multiple experiments.
  • - An experiment using N-decane validated the new calculation method, showing a small error range (-0.61% to 1.24%) in temperature estimates between 295.50-609.38 K, indicating its effectiveness for fluid temperature estimation in calorimetric applications.

Article Abstract

The distributed flow calorimeter is employed to measure reaction enthalpy, mixing enthalpy, and other thermodynamic parameters of hydrocarbon fuels. This information serves as a foundation for selecting and developing hydrocarbon fuels for hypersonic flight. The fluid temperature in the tube is a key factor in characterizing its thermodynamic behavior. Given the challenges of monitoring fluid temperature within a tube using current flow calorimetry, a distributed method for calculating fluid temperature in tubes under high-temperature conditions is proposed. This method realizes the interpolation of the enthalpy function of the experimental fluid through several sets of experiments with varying power levels. The fluid temperature in the tube is calculated by considering the microelement as the research object. First, the methodology for calculating fluid temperature in narrow pipes across a wide temperature range is presented. Second, the simulation model of the flow calorimeter is established, and the methodology is verified through numerical simulation. Finally, a flow calorimetric experimental device is setup. N-decane was used as the fluid in the experiment, and the temperature was calculated, and the calculated results were compared with the NIST data. In the temperature range of 295.50-609.38 K, the relative error range of the calculation of n-decane temperature is -0.61% to 1.24%. The experimental results show that the method can effectively estimate the fluid temperature of the distributed flow calorimeter.

Download full-text PDF

Source
http://dx.doi.org/10.1063/5.0231588DOI Listing

Publication Analysis

Top Keywords

fluid temperature
28
flow calorimeter
16
temperature range
12
distributed flow
12
temperature tube
12
temperature
11
fluid
9
wide temperature
8
hydrocarbon fuels
8
calculating fluid
8

Similar Publications

Good fat vs bad fat in Milk: A molecular level Understanding of Indian cow milk using confocal Raman microscopy.

Spectrochim Acta A Mol Biomol Spectrosc

January 2025

Department of Chemistry, Institute of Science, Banaras Hindu University, Varanasi 221005, India. Electronic address:

Milk, a complex fluid renowned for abundance of vitamins and immune-boosting antibodies, holds a pivotal position in human nutrition. The research delves into the fundamental constituents of milk, focusing on cis-fatty acids (cis-FA), trans-fatty acids (trans-FA), and theα-helixstructure found in proteins. These constituents are instrumental in the determination of milk quality and its nutritional value.

View Article and Find Full Text PDF

Menstrual cycle effects on thermoregulation while exercising in the heat.

J Therm Biol

January 2025

School of Integrative Physiology and Athletic Training, University of Montana, 32 Campus Drive, Missoula, MT, 59812, USA. Electronic address:

Women may be challenged to maintain thermoregulation due to hormonal changes associated with the menstrual cycle. The purpose of this study was to assess the effect of the menstrual cycle phase on core temperature, hydration status, and perceived exertion while exercising under uncompensable heat gain. Eleven eumenorrheic women (24.

View Article and Find Full Text PDF

Advanced adiabatic compressed air energy storage systems dynamic modelling: Impact of the heat storage device.

Heliyon

January 2025

IFP Energies nouvelles, 1 et 4 avenue de Bois Préau, 92852, Rueil-Malmaison, France.

Advanced Adiabatic Compressed Air Energy Storage (AACAES) is a technology for storing energy in thermomechanical form. This technology involves several equipment such as compressors, turbines, heat storage capacities, air coolers, caverns, etc. During charging or discharging, the heat storage and especially the cavern will induce transient behavior of operating points, notably temperature, pressure, and volume flow.

View Article and Find Full Text PDF

Background: Biomarkers are essential tools for diagnosing diseases. Saliva, as a human fluid, effectively reflects the body's condition due to its rich composition. Analyzing saliva components allows for noninvasive, cost-effective, and time-efficient screening and diagnosis.

View Article and Find Full Text PDF

Orodispersible film (ODF) is one of the novel formulations that disintegrate rapidly in the mouth without the requisite for water compared to other conventional oral solid dosage formulations. This delivery system serves as a convenient mode of administration, especially in patients who have dysphagia and fluid restriction, being beneficial to pediatric, geriatric, and bedridden patients. A novel sildenafil ODF containing sildenafil citrate is formulated to be used in patients with erectile dysfunction (ED).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!