Comparison of grooved and conventional Macintosh laryngoscope blades for throat pack insertion: A randomized control trial.

J Anaesthesiol Clin Pharmacol

Department of Anaesthesiology, Kasturba Medical College, Manipal, Manipal Academy of Higher Education Manipal, Karnataka, India.

Published: March 2024

AI Article Synopsis

  • A study was conducted to evaluate a new design of the Macintosh laryngoscope blade with a groove to support the endotracheal tube during throat pack insertion in otolaryngological surgeries.
  • The trial involved 80 patients, comparing the effectiveness of the grooved blade (group G) to the conventional blade (group C) in terms of tube stability, ease of packing, and time taken.
  • Results showed that the grooved blade kept the tube centered in 35 patients versus 10 in the conventional group, allowed for easier packing, and reduced the time needed for the throat pack insertion.

Article Abstract

Background And Aims: Throat pack insertion is required for many otolaryngological surgeries under general anesthesia. To facilitate surgery, an oral Ring, Adair, and Elwyn (RAE) endotracheal tube is used. When the Macintosh blade is used to insert the pack, the tube often moves to the left as per its design. We designed a groove in the Macintosh laryngoscope blade to hold the endotracheal tube in place during pack placement. This study was done to compare the grooved and conventional Macintosh laryngoscope blades for throat pack insertion with respect to the intraoral movement of the tube, ease of packing, time taken, and associated hemodynamic changes in patients who have been intubated with a preformed oral endotracheal tube.

Material And Methods: This randomized control trial included 80 patients after approval from the institutional ethics committee and CTRI. Informed written consent was taken from all patients. Packing of cases was done with either grooved (group G) or conventional Macintosh laryngoscope blade (group C). Student's -test, Chi-square test, and Mann-Whitney U test were used to analyze data.

Results: The tube remained in the center in 35/40 patients in group G as compared to 10/40 patients in group C ( < 0.0001). Packing the throat was significantly easier ( = 0.005), and time taken to pack the throat was shorter in group G as compared to group C ( = 0.0008).

Conclusions: The endotracheal tube remains in the center more often during throat pack insertion using a grooved laryngoscope blade as compared to the conventional Macintosh laryngoscope blade. The grooved blade is easier to use and takes less time to insert the throat pack.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11463916PMC
http://dx.doi.org/10.4103/joacp.joacp_318_23DOI Listing

Publication Analysis

Top Keywords

macintosh laryngoscope
20
throat pack
20
conventional macintosh
16
pack insertion
16
laryngoscope blade
16
endotracheal tube
12
grooved conventional
8
laryngoscope blades
8
blades throat
8
pack
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!