Does sheath size matter: benchtop comparison of flow and pressure across variety of continuous flow endoscopes.

World J Urol

Department of Urology, University of Minnesota, 420 Delaware St. Se. MMC 394, Minneapolis, MN, 55455, USA.

Published: October 2024

AI Article Synopsis

  • The study evaluated how different endoscope and sheath sizes affect irrigation outflow and bladder pressure during laser enucleation procedures.
  • Using a silicone bladder model, five sheath combinations were tested, revealing that the Storz 28/26Fr had the highest outflow rate and lowest bladder pressure.
  • Results indicate that larger sheath sizes improve outflow, while smaller sizes lead to increased bladder pressure, suggesting a need for careful sheath selection to enhance surgical outcomes.

Article Abstract

Introduction: Laser enucleation utilizes purpose-built endoscopes for laser stabilization and continuous flow. No evaluation has been done with respect to flow or intravesical pressure with these scopes. We sought to evaluate the effect different endoscopes and sheath sizes on irrigation outflow and intravesical pressure.

Methods: Using a benchtop model using a silicone bladder model, five outer/inner sheath combinations were assessed: Storz 28/26Fr, Storz 26/26Fr, Wolf 26/24Fr, Wolf 26/22Fr, and Wolf 24/22Fr. A urodynamics pressure transducer was inserted alongside the scope for bladder pressure measurement and outflow from scope to drain was measured using uroflowmetry device. Four 1-minute trials were recorded for each sheath and the steady state flow and pressure was recorded.

Results: The Storz 28 F outer sheath and 26 F inner sheath had the highest outflow (12.4 ± 0.5 mL/s, p < 0.01). The Wolf 24 F outer and 22 F inner had the lowest outflow (7.0 ± 0.0 mL/s, p < 0.01). The steady state bladder pressure was the lowest in the Storz 28/26 (1.5 ± 1.7 cm H2O, p < 0.01)) and the greatest in the Storz 26/26 (24.2 ± 1.9 cm H2O, p < 0.01).

Conclusion: The Storz 28/26 combination had best outflow rate and lowest intravesical pressures in our benchtop study. Flow rates generally decreased with smaller sheath sizes and steady state bladder pressures increased as the difference between the outflow and inflow sheath size narrowed. These findings provide initial parameters that could guide sheath selection in future to optimize visualization and success of voiding trials.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00345-024-05270-0DOI Listing

Publication Analysis

Top Keywords

flow pressure
8
continuous flow
8
sheath
6
flow
5
pressure
5
sheath size
4
size matter
4
matter benchtop
4
benchtop comparison
4
comparison flow
4

Similar Publications

Objectives: The effects of acute physical exercise in patients with resistant hypertension remain largely unexplored compared with hypertensive patients in general. We assessed the short-term effects of acute moderate-intensity (MICE) and high-intensity interval exercise (HIIE) on the clinic (BP) and 24-h ambulatory blood pressure (ABP) of patients with resistant hypertension.

Methods: Using a crossover randomized controlled design, 10 participants (56 ± 7 years) with resistant hypertension performed three experimental sessions: MICE, HIIE, and control.

View Article and Find Full Text PDF

While soundscapes shape the structure and function of auditory systems over evolutionary timescales, there is limited information regarding the adaptation of wild fish populations to their natural acoustic environments. This is particularly relevant for freshwater ecosystems, which are extremely diverse and face escalating pressures from human activities and associated noise pollution. The Siamese fighting fish is one of the most important cultured species in the global ornamental fish market and is increasingly recognized as a model organism for genetics and behavioural studies.

View Article and Find Full Text PDF

Balancing pH and Pressure Allows Boosting Voltage and Power Density for a H-I Redox Flow Battery.

ACS Appl Energy Mater

January 2025

Department of Chemical Engineering, Delft University of Technology, Van der Maasweg 9, 2629HZ Delft, The Netherlands.

The decoupled power and energy output of a redox flow battery (RFB) offers a key advantage in long-duration energy storage, crucial for a successful energy transition. Iodide/iodine and hydrogen/water, owing to their fast reaction kinetics, benign nature, and high solubility, provide promising battery chemistry. However, H-I RFBs suffer from low open circuit potentials, iodine crossover, and their multiphase nature.

View Article and Find Full Text PDF

Purpose: Despite significant improvements in the design and performance of continuous flow left ventricular assist devices (CFLVADs), one of the most important reasons hampering further penetration of this technology is the occurrence of adverse events, especially strokes. One of the well-known risk factors for strokes is hypertension which is particularly common in patients undergoing a CFLVAD implant. While the device is implanted in the heart, strokes happen due to pathology in the brain and we hypothesised that modelling the blood flow in the circle of Willis might shed light on the causation of strokes in this situation.

View Article and Find Full Text PDF

Purpose: To assess platelet-rich plasma (PRP) changes in platelet and leucocyte count, insulin-like growth factor 1 (IGF-1), and interleukin 6 (IL-6) concentration after bilateral low-load knee extensions under blood flow restriction (BFR).

Methods: The present randomised controlled trial protocol will include two groups: the intervention group, which will undergo bilateral knee extensions under BFR, and the control group, which will perform bilateral knee extensions without BFR. Participants will be randomly allocated in a 1:1 ratio.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!