Background: Despite adherence to reprocessing protocols, duodenoscopes frequently remain contaminated, highlighting significant knowledge gaps in reprocessing efficiency.
Aim: To identify risk factors in duodenoscope reprocessing procedures affecting contamination rates.
Methods: Cultures from Pentax ED34-i10T2 duodenoscopes collected between February 2022 and December 2023 were included. Contamination was determined by the presence of micro-organisms of gut or oral origin (MGO). Data on duodenoscope use, reprocessing lead times and personnel were retrieved from electronic medical records. Risk factors were derived from reprocessing guidelines and literature. These included a delay >30 min in initiating manual cleaning, manual cleaning duration of ≤5 min, drying time <90 min, personnel reprocessing frequency, and storage exceeding seven days. A logistic mixed-effects model evaluated these factors' impact on duodenoscope contamination.
Findings: Out of 307 duodenoscope cultures, 58 (18.9%) were contaminated with MGO. Throughout the study period, the duodenoscopes underwent 1296 reprocessing cycles. Manual cleaning times of ≤5 min significantly increased contamination odds (adjusted odds ratio (aOR): 1.61; 95% confidence interval (CI): 1.10-2.34; P = 0.01). Increased usage of a duodenoscope was associated with reduced odds of contamination (aOR: 0.80; 95% CI: 0.64-0.995; P = 0.045). Other studied risks showed no clear association with contamination rates.
Conclusion: Manual cleaning times of ≤5 min increased the odds of contamination with MGO. Delays in reprocessing initiation and incomplete drying, traditionally considered as risk factors, were not associated with an increased risk of contamination in this study. Future research should explore whether enhanced surveillance of reprocessing times can mitigate duodenoscope contamination.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jhin.2024.09.018 | DOI Listing |
J Gastroenterol Hepatol
November 2024
Division of Gastroenterology, Department of Medicine, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand.
Background And Aim: To reduce bacterial contamination after reprocessing, various new designs of duodenoscopes have been developed to better expose the elevator complex for cleaning. We compared the rates of bacterial contamination and organic residue in disposable distal cap duodenoscopes and detachable elevator duodenoscopes after manual cleaning and high-level disinfection (HLD), as well as their cost-effectiveness.
Methods: A total of 162 duodenoscopes were randomly assigned to either Group A (disposable distal caps; n = 81) or Group B (detachable elevator; n = 81).
J Hosp Infect
December 2024
Department of Gastroenterology and Hepatology, Erasmus MC University Medical Center, Rotterdam, The Netherlands. Electronic address:
Background: Despite adherence to reprocessing protocols, duodenoscopes frequently remain contaminated, highlighting significant knowledge gaps in reprocessing efficiency.
Aim: To identify risk factors in duodenoscope reprocessing procedures affecting contamination rates.
Methods: Cultures from Pentax ED34-i10T2 duodenoscopes collected between February 2022 and December 2023 were included.
Am J Gastroenterol
September 2024
Goethe University Frankfurt, University Hospital, Medical Clinic 1, Germany.
J Clin Gastroenterol
December 2024
Department of Gastroenterology and Hepatology, Mount Sinai South Nassau, Oceanside, NY.
Endoscopic retrograde cholangiopancreatography (ERCP) is the gold standard in the endoscopic management of biliary disease. An average of 700,000 ERCPs are performed every year, and most are performed using a reusable flexible duodenoscope. The innovation of disposable duodenoscopes has changed the dynamic in the advanced endoscopy field of study to primarily reduce or eliminate the risk of cross-contamination between patients.
View Article and Find Full Text PDFInfect Control Hosp Epidemiol
March 2024
Division of Infectious Diseases, Henry Ford Health, Detroit, MI, USA.
Background: Whole-genome sequencing (WGS) has emerged as an alternative genotyping tool for outbreak investigations in the healthcare setting. We describe the investigation and control of a New Delhi metallo-B-lactamase (NDM)-producing cluster in Southeast Michigan.
Methods: Michigan Bureau of Laboratories identified several closely related NDM-producing isolates with WGS.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!