Article Abstract

Background: Major efforts in computed tomography (CT) have been focused on reducing radiation dose to patients while maintaining adequate diagnostic quality. To that end, research tools have been developed to simulate reduced-dose images via either image-based or projection-based methods. The former is limited to fully capturing realistic texture, streak, and non-stationary characteristics of reduced dose, while the latter is impractical clinically.

Purpose: To develop and validate an image-based noise addition method that accounts for such attributes while being practical in clinical settings.

Methods: A noise addition method was developed to add realistic noise in the image domain. The method first estimates the noise power spectrum (NPS) of CT images, which are also forward-projected to form synthetic projections. The projection data are supplemented with random white noise proportional to their attenuation values. The noise sinogram is then back-projected onto the image, filtered by the NPS, and scaled according to the desired dose reduction level. The tool was evaluated using both phantom images and patient data. The phantom images were acquired using a multi-sized image quality phantom (Mercury Phantom 3.0, Duke University), and a thorax anthropomorphic phantom (Lungman Phantom, Kyoto Kagaku) at different dose levels and reconstruction settings. The patient images consisted of two dose levels of various CT examinations and reconstruction settings. The simulated and real reduced-dose images were compared in terms of the noise magnitude and texture (i.e., NPS average frequency, NPS-f). The utility of this methodology was also assessed for routine clinical use for CT protocol review.

Results: For the phantom images, the percent errors in the noise magnitude between the simulated images and the actual images of the Mercury Phantom and anthropomorphic phantom images were 3.34% and 3.50%, respectively. The difference in f was 0.07 mm for the Mercury Phantom and 0.06 mm for the anthropomorphic phantom between the simulated and actual images. The average noise magnitude percent error between the simulated and actual patient images was 4.61% with noise texture judged to be visually comparable with some kernel dependencies. When implemented clinically, the tool proved practical to simplify the process of estimating radiation dose reduction for CT protocols, resulting in a 50% dose reduction of our multiple myeloma protocol.

Conclusions: The method generated simulated CT images with realistic noise properties similar to images acquired at the same radiation exposure without needing access to raw projection data.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11700769PMC
http://dx.doi.org/10.1002/mp.17444DOI Listing

Publication Analysis

Top Keywords

phantom images
16
images
15
noise
12
noise addition
12
addition method
12
dose reduction
12
mercury phantom
12
anthropomorphic phantom
12
noise magnitude
12
phantom
11