  • The study evaluates the safety and effectiveness of the Inno-Xmart braided venous stent system for treating iliofemoral venous obstruction in patients exhibiting symptoms.
  • The findings indicate a high surgical success rate of 95.3% and a low major adverse event rate of 5.1% over 12 months, with no stent fractures reported.
  • Patients also experienced significant improvements in quality of life measures, showcasing the stent’s positive impact on both medical outcomes and patient well-being.

Article Abstract

This study aims to evaluate the safety and efficacy of the dedicated Inno-Xmart braided venous stent system (Suzhou Innomed Medical Device Co., Ltd., Jiangsu, China) in treating symptomatic iliofemoral venous obstruction. This clinical study followed a prospective, multicentre, single-arm design with the application of an objective performance goal. Patients diagnosed with symptomatic iliofemoral venous obstruction who met the eligibility criteria of this study were enrolled and treated with the Inno-Xmart venous stent system. The safety endpoints included the assessment of stent fracture, satisfaction of delivery system and 12-month incidence rate of major adverse events (MAEs). The primary efficacy endpoint focused on evaluating the 12-month primary patency rate through venography as determined by core laboratory. Secondary efficacy endpoints included surgical success rate, 6-month primary patency rate and the changes in quality of life from baseline to 6- and 12-month follow-up intervals. Between September 18, 2019, and April 26, 2021, 193 patients were successfully enrolled across 18 research institutions. The surgical success rate was 95.3% (184/193), the 12-month MAE rate was 5.1% (9/178) with no stent fractures or migrations. The 12-month primary patency rate for the participants was 96.1%, significantly surpassing the literature-derived objective performance of 80% (95% confidence interval [CI], 92.1-98.4; P < 0.0001). In addition, the mean venous clinical severity score (VCSS) and Chronic Venous Disease Quality of Life Questionnaire (CIVIQ) scores at the 6- and 12-month follow-ups were significantly lower than the preoperative scores (P < 0.001). The innovative, dedicated braided venous stent designed to address symptomatic iliofemoral venous obstruction demonstrates a high technical success rate, low complication rates, and impressive mid-term (12-month) patency. It effectively enhanced the quality of life for patients and holds promising prospects for a wide range of applications. The clinical study was officially registered in the "Chinese Clinical Trial Registry" (Registration number: ChiCTR2000040216, date of registration: November 25th, 2020).

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12928-024-01037-xDOI Listing

Publication Analysis

Top Keywords

venous stent
12
stent system
12
symptomatic iliofemoral
12
iliofemoral venous
12
venous obstruction
12
primary patency
12
patency rate
12
safety efficacy
8
inno-xmart venous
8
objective performance
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!