Background: The Rapid-O2 oxygen insufflation device® (Rapid-O2) was designed primarily for rescue oxygenation in cannot intubate, cannot oxygenate (CICO) events; thus, hypercapnia is inevitable. Rapid-O2 was modified to enhance ventilation using the Venturi effect during expiration.
Methods: To determine the most effective combination of inner catheters (20 G, 18 G, 16 G, 14 G, and 2-mm inner diameter [ID] transtracheal catheter [TTC]) and insufflation catheters (16 G, 14 G, and 2-mm ID TTC) for achieving optimum ventilation, insufflating and expiratory flows were measured at an oxygen flow rate of 15 L/min. The insufflating and expiratory pressures were measured at 6-15 L/min. The flows and pressures were measured using a gas flow analyzer. The insufflating and expiratory times were measured using a trachea-lung model to obtain minute volumes. To assess the improvement by modifying the Rapid-O2, minute volumes were measured using the Rapid-O2.
Results: The most appropriate inner catheter was 18 G. The insufflating pressures ranged from 97 (2-mm ID TTC) to 377 cmH2O (16 G) at 15 L/min. During expiration, similar negative pressures of 50 cmH2O were measured in the insufflation catheters at 15 L/min. At lung compliance of 100 ml/cmH2O, the minute volumes through a 2-mm ID and 14 G insufflation catheters were 7.0 and 5.37 L/min, respectively, at 15 L/min. The minute volumes were significantly greater in modified Rapid-O2.
Conclusions: Modified Rapid-O2 provided sufficient minute volumes in adults using a 14 G or 2-mm ID insufflation catheter at 15 L/min, demonstrating its potential for ventilation in CICO events.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4097/kja.24095 | DOI Listing |
Eur J Nucl Med Mol Imaging
December 2024
Department of Nuclear Medicine, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 18, Bern, 3010, Switzerland.
Purpose: Long axial field-of-view (LAFOV) positron emission tomography/computed tomography (PET/CT) scanners enable high sensitivity and wide anatomical coverage. Therefore, they seem ideal to perform post-selective internal radiation therapy (SIRT) Y scans, which are needed, to confirm that the dose is delivered to the tumors and that healthy organs are spared. However, it is unclear to what extent the use of LAFOV PET is feasible and which dosimetry approaches results in accurate measurements.
View Article and Find Full Text PDFPharmacy (Basel)
December 2024
Department of Clinical and Applied Movement Sciences, Brooks College of Health, University of North Florida, Jacksonville, FL 32224, USA.
While previous research has linked physical activity (PA) with lower prescription medication consumption, limited evidence has investigated sedentary time (ST) as a major contributor to medication use, despite ST's known association with chronic disease and mortality risk, even when PA volume is considered. This study aimed to examine the independent associations between objectively measured ST, patterns of sedentary bouts, and self-reported prescription medication use among adults ≥25 years of age. Thirty-two participants reported the number and type of medications they were currently prescribed and wore an accelerometer continuously on their hip for seven days to detect their ST.
View Article and Find Full Text PDFBiosensors (Basel)
December 2024
Department of Smart Health Science and Technology, Kangwon National University, Chuncheon 24341, Republic of Korea.
Microfluidic devices have revolutionized biosensing by enabling precise manipulation of minute fluid volumes across diverse applications. This review investigates the incorporation of machine learning (ML) into the design, fabrication, and application of microfluidic biosensors, emphasizing how ML algorithms enhance performance by improving design accuracy, operational efficiency, and the management of complex diagnostic datasets. Integrating microfluidics with ML has fostered intelligent systems capable of automating experimental workflows, enabling real-time data analysis, and supporting informed decision-making.
View Article and Find Full Text PDFJMIR Cardio
December 2024
New South Wales Ambulance, New South Wales Health, Sydney Olympic Park, Australia.
Background: Emergency medical services attend out-of-hospital cardiac arrests all across Australia. Resuscitation by emergency medical services is attempted in nearly half of all cases. However, resuscitation skills can degrade over time without adequate exposure, which negatively impacts patient survival.
View Article and Find Full Text PDFPhysiol Rep
December 2024
Integrative Physiology Laboratory, University of Illinois at Chicago, Chicago, Illinois, USA.
Multiple sclerosis (MS) is a chronic neurological condition resulting in decreased aerobic capacity (peak VO). The hemodynamic responses to peak exercise in MS are unknown. Further, it is unknown if the hemodynamic responses are due to disease or fitness.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!