Objectives: Previous finite element analyses (FEA) have shown promising results for using two titanium screws in treating mandibular condylar head fractures but limited mechanical stability of a two-screw osteosynthesis with magnesium screws. Given the potential benefits of magnesium screws in terms of biocompatibility and resorption, this study aimed to compare two- and three-screw osteosynthesis solutions for a right condylar head fracture (AO CMF type p) with magnesium screws with a FEA.
Materials And Methods: A previously validated finite element model simulating a 350 N bite on the contralateral molars was used to analyze von Mises stress within the screws, fragment deformation, and fracture displacement. All screws were modeled with uniform geometric specifications mirroring the design of Medartis MODUS Mandible Hexadrive cortical screws.
Results: The three-screw configuration demonstrated lower values for all three parameters compared to the two-screw scenario. There was a 30% reduction in maximum von Mises stress for the top screw and a 46% reduction for the bottom screw.
Conclusions: Fracture treatment with three magnesium screws could be a valuable and sufficiently stable alternative to the established treatment with titanium screws. Further studies on screw geometry could help improve material stability under mechanical loading, enhancing the performance of magnesium screws in clinical applications.
Clinical Relevance: The use of magnesium screws for osteosynthesis of mandibular condylar head fractures offers the benefit of reducing the need for second surgery for hardware removal. Clinical data is needed to determine whether the advantages of resorbable screw materials outweigh potential drawbacks in condylar head fracture treatment.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC11427473 | PMC |
http://dx.doi.org/10.1007/s00784-024-05927-5 | DOI Listing |
Case Rep Dent
December 2024
Department of Prosthodontics, Geriatric Dentistry and Craniomandibular Disorder, Charité-Universitätsmedizin Berlin, Aßmannshauser Straße 4-6 14197, Berlin, Germany.
Due to bone loss, implant placement in the posterior mandible is often impossible without prior augentative procedures. The reconstruction of bone defects with horizontal and vertical components using particulated bone grafts requires the placement of a mechanically stable structure for stabilization of the grafting material. Although titanium-reinforced membranes and titanium meshes have been shown to be effective in this indication, the necessity of their removal, often in a separate surgical procedure, is seen as a disadvantage.
View Article and Find Full Text PDFMaxillofac Plast Reconstr Surg
December 2024
Department of Oral and Maxillofacial Surgery, Hallym University Sacred Heart Hospital, Anyang, 14066, Republic of Korea.
Background: Titanium (Ti-6Al-4 V) is used for fixation in LeFort I osteotomy, a procedure for treating midface deformities. This study assessed the biomechanical stabilities of two Mg alloys (WE43 and ZK60) as biodegradable alternatives and compared them against Ti using finite element analyses. The LeFort I osteotomy procedure was simulated, and various plate and screw configurations were tested.
View Article and Find Full Text PDFInt J Pharm
December 2024
Chemical and Biochemical Engineering, Rutgers University, 98 Brett Road, Piscataway, NJ 08854, USA. Electronic address:
Favipiravir, a high dose antiviral drug effective for oral treatment for COVID-19, with poor water solubility is formulated using a simple, low-cost melt coating and granulation methodology. High-dose (82.5 % w/w API) tablets (600 mg and 800 mg) with desired release profiles are developed while minimizing excipient burden.
View Article and Find Full Text PDFMaterials (Basel)
November 2024
Department of Orthopedics Surgery, Jeonbuk National University Hospital, Jeonju 54907, Republic of Korea.
Biodegradable material, such as magnesium alloy or polylactic acid (PLA), is a promising candidate for orthopedic surgery. The alloying of metals and the addition of rare earths to increase mechanical strength are still questionable in terms of biosafety as absorbent materials. Therefore, the purpose of this study is to understand the effect of substances due to the degradation of various biodegradable substances on organs in the body or surrounding tissues.
View Article and Find Full Text PDFBioact Mater
February 2025
Laboratory of Metal Physics and Technology, Department of Materials, ETH Zurich, 8093 Zurich, Switzerland.
Magnesium alloys present a compelling prospect for absorbable implant materials in orthopedic and trauma surgery. This study evaluates an ultra-high purity, lean magnesium-calcium alloy (X0), both with and without plasma electrolytic oxidation (PEO) surface modification, in comparison to a clinically utilized WE43 magnesium alloy. It is shown that the mechanical properties of X0 can be tuned to yield a high-strength material suitable for bone screws (with an ultimate tensile strength of 336 MPa) or a ductile material appropriate for intraoperatively deformable plates (with an elongation at fracture of 24 %).